244 actuator: maintenance, 298 hsm: maintenance, 244 actuator: maintenance 86 – MTS Series 322 Load Frames User Manual

Page 86: 298 hsm: maintenance 86

Advertising
background image

Series 322 Load Unit

86

Maintenance

244 Actuator: Maintenance

The series 244 Actuator is designed for extended periods of operation without
extensive maintenance requirements.

A summary of the routine maintenance procedures is listed below:

Weekly

Clean exposed areas of the actuator piston rod with a clean, dry, lint free rag. If
the actuator is continually exposed to a dirty operating environment, clean the
piston rod on a daily basis.

Monthly

Inspect actuator piston rod and seals for excessive wear and/or leakage. Small
scratches in the axial direction of the piston rod or polishing of the rod surface is
considered normal operating wear.

Yearly

Change actuator seals if necessary. Actuator assemblies can require more or less
frequent seal changes depending on usage. External oil leakage and/or decreased
performance are indicators of seal wear. Seal replacement is considered a service
procedure; contact you MTS Service Engineer for additional information.

298 HSM: Maintenance

The series 298 Hydraulic Service Manifold requires the filter be changed
periodically.

Component Identification

Pressure

Accumulator

Return

Accumulator

Inlet

Filter

Dirty

Filter

Indicator

Advertising