Steps to Run a Test

MTS Series 793 Controller Overview

Perform a Practice Test

91

E.

Install a dummy specimen that has a compliance similar to the test
specimen.

•

From the Station Controls panel toolbar, click the Manual Command

icon .

•

Click to select the Enable Manual Command check box.

•

Select the Displacement control mode.

•

Adjust the Manual Command slider to position the actuator while
installing the specimen.

•

Switch to the Force control mode.

F.

Adjust the P Gain and D Gain tuning controls.

•

Start the Function Generator by clicking the Program Run control
and observe the resulting waveform on the Scope.

•

From the Display menu, select Station Setup.

•

From the tree view, expand Channels, then select Channel 1.

•

Click the Tuning icon

to display the tuning adjustment controls.

•

Increase the P Gain control until you see a little overshoot and a little
ringing on the Scope.

•

Slowly increase the D Gain control to reduce the overshoot and
ringing. Keep in mind that small adjustments of D can have large
effects on your system.

•

Repeat P Gain and D Gain adjustments until you achieve an optimum
waveform

Ringing

Overshoot