5. Reset the interlocks if needed and select the Low, then High, power buttons.

6. Click Run to run the test.

5.4.2.0 Set the FGC DCPD Test Parameters

5.4.2.1.0 Test Parameters Overview

After a new test run is added to the test, the Main Menu window appears. This window provides access to
all necessary test parameters. Click the parameter buttons in the Main Menu window to specify parameter
values for the test.

5.4.2.2.0 FCG DCPD Parameter Values for Simulation

The sample test includes a test run with test data that can be analyzed. If you want to reproduce the same
results in another test run, add another test run, and then specify the following values for the test parameters.
Other parameters not listed here should remain unchanged.

Precrack Parameters

Precrack Parameter Values

Value to Use for Simulation

Parameter

10 Hz

Precrack Frequency

25 mm

Precrack Final Crack Limit

1 kN

Precrack Measure Load Level

0.1 unitless

Precrack Load Ratio

0.1 kN/mm^1.5

Precrack Final Maximum K

1000000 count

Precrack Cycle Limit

Crack Growth Parameters

Delta K Parameter Values

Value to Use for Simulation

Parameter

True Sine

FCG Wave Shape

10 Hz

FCG Frequency

0.2 kN/mm^1.5

FCG Delta-K Initial

0.1 unitless

FCG Load Ratio

MTS TestSuite 57