Precrack Parameters

Description

Parameter

Specify the desired crack size to achieve in
the precrack activity. When the specified
crack size is achieved, the activity stops.

Precrack Final Crack
Limit

Specify the cycle frequency of the command
signal for the precrack activity.

Precrack Frequency

Specify the ratio of the minimum to maximum
load applied to the specimen. The minimum

Precrack Load Ratio

load is determined by this value and the
maximum load is specified by you.

Specify the percentage of the measured load
range used as the lower limit for linear least

Precrack Lower Least
Squares Fit Percentage

squares fit compliance determination. This
parameter is not used for FCG DCPD tests.

Specify the percentage of the measured load
range used as the upper limit for linear least

Precrack Upper Least
Squares Fit Percentage

squares fit compliance determination. This
parameter is not used for FCG DCPD tests.

Specify the load level at which the voltage is
measured. The load is applied at this level

Precrack Measure Load
Level

to ensure the crack is open. This parameter
is used only for FCG DCPD tests.

Specify the number of cycles at which
precrack activity stops.

Precrack Cycle Limit

Specify the maximum stress intensity desired
at the end of precrack activity. The initial

Precrack Final Maximum
K

precrack Maximum K is 1.4 times the Final
Maximum K value.

5.5.2.4.0 JIC Parameters

In the JIC Parameters section of the main menu, you select COD (Crack Opening Displacement) or
Displacement as the control mode for the test, and then specify the parameters for the selected mode.

74 MTS TestSuite

5.0 Running Simulation Tests

Advertising