5.5.2.5.0 COD Mode Parameters

COD Mode Parameters

Description

Parameter

Specify the amount of force, in force units,
to unload in both the unload and reload
segments.

JIC Absolute Unload

Specify the COD increment that will be
added to the current COD level at the

JIC COD Increment

beginning of each ramp segment to
determine a maximum COD for this step. If
the maximum COD is reached before the
maximum displacement or load level is
reached, it triggers the next segment of the
step.

Specify the length of time to hold at constant
COD or Displacement in the hold crack

JIC Crack Propagation Hold
Time

propagation activity. If this time is zero, the
hold crack propagation activity is bypassed.

Specify the inverse of the compliance of the
test system. This term is used to determine

JIC Frame Stiffness

the displacement due to load frame
deflection and the deformation of fixtures.
The resulting displacement is subtracted
from the measured displacement when
calculating the energy absorbed by the
specimen.This is particularly important when
using SE(B) specimens.

Specify the ramp rate for the initial ramp
segment when it is done in load control.

JIC Load Ramp Rate

Specify the number of unload and reload
segment pairs to perform in each step
increment of the JIC test.

JIC Number of Unloads

Specify the percentage of the load of the final
precrack cycle to use as the load for the end
of the first ramp segment.

JIC Percent of Final
Precrack Load

Specify the amount of force to unload as a
percentage of the increment peak load.

JIC Percent Unload

Specify a value other than zero to apply a
displacement limit during the ramp segment

JIC Ramp Displacement
Limit Increment

of a JIC step. The limit trips if the measured

MTS TestSuite 75

5.0 Running Simulation Tests