Power failure mode, Low noise mode, Audible alarm enable – New Era NE-1000 Multi-Phaser Programmable Syringe User Manual

Page 20: Ttl i/o operational trigger default configuration

Advertising
background image

New Era Pump Systems Inc.

www.SyringePump.com

Model NE-1000 Multi-Phaser™

Publication #1200-01

15

08/01/14

nn

RS-232 pump network address, where ‘nn’ is the network address.

nnnn

RS-232 pump network baud rate, where ‘nnnn’ indicates the baud rate.

Otherwise, the current communications mode will be displayed:

,

, or

Reciprocating, Dual Pump, or Alternating

Communications Mode.

8.1

Power Failure Mode

n

Setting: ‘0’ = Disabled, ‘1’ = Enabled.

When enabled, if the Pumping Program was operating when power to the pump was disrupted, the Pumping
Program will automatically start operating when power is reconnected to the pump. Pressing any key on the
keypad while powering up the pump will stop the Pumping Program from starting.

CAUTION: The Pumping Program will start operating from the beginning of the Pumping Program

(Phase 1), regardless of what part of the Pumping Program was operating when the power was disrupted.

When the NE-1000 syringe pump is used as a component in an automated infusion/withdrawal dispensing
system, a Pumping Program can be designed to automatically synchronize the pusher block at the start of the
Pumping Program. This would be accomplished using attached sensors that send signals to the Pumping
Program.

8.2

Low Noise Mode

n

Setting: ‘0’ = Disabled, ‘1’ = Enabled.

A side effect of the NE-1000’s high precision micro-stepped motor driver is a high frequency resonance sound
at very low pumping speeds. This mode minimizes this sound by reducing the micro-stepping, increasing
pulsations.

8.3

Audible Alarm Enable

n

Setting: ‘0’ = Disabled, ‘1’ = Enabled.

When alarms are enabled, a steady buzzer alarm will sound during alarm conditions, such as when the motor
stalls: Pressing any key will stop the alarm.

8.4

TTL I/O Operational Trigger Default Configuration

aa

Configures the default mode of the TTL I/O ‘Operational Trigger’ (pin 2) that will control the

Pumping Program’s operation. This default setting can be overridden within a Pumping Program. (See sec.
11.1, TTL I/O Operational Controls). The 2 letter configuration Setting parameter to the right of the colon (:)
is defined in column ‘Setting’ as follows:

Setting

<aa>

RS-232 Command Setting

FUN TRG <n>

<n> =>

Name

Function

Ft

0

Foot Switch

Falling edge starts or stops the Pumping Program

FH

1

Foot Switch Hold

Falling edge starts the Pumping Program
Rising edge stops the Pumping Program

F2

2

Foot Switch Reversed

Rising edge starts or stops the Pumping Program

LE

3

Level Control

Falling edge stops the Pumping Program
Rising edge starts the Pumping Program

St

4

Start Only

Falling edge starts the Pumping Program

t2

5

Start Only Reversed

Rising edge starts the Pumping Program

SP

6

Stop Only

Falling edge stops the Pumping Program