New Era NE-1000 Multi-Phaser Programmable Syringe User Manual

Page 45

Advertising
background image

New Era Pump Systems Inc.

www.SyringePump.com

Model NE-1000 Multi-Phaser™

Publication #1200-01

40

08/01/14

‘Volume’ and Pumping Direction can be changed. Cannot be enabled when the Pumping Program is
currently programmed with a multiple Phase Program.

SET KEYPAD AND NOTIFICATION BEEP
BP [ <on-off> ]

Set/query keypad and notification beep mode. Set beep enables or disables beep mode.

10.4.4

General Control and Status Commands

TTL I/O OUTPUT SETTING
OUT <n> <TTL level>

Sets TTL level on user definable output pin on the ‘TTL I/O’ connector.

<n> Indicates pin number on ‘TTL I/O’ connector

Valid value: 5 (Program Output pin)

Query is not applicable.

TTL INPUT QUERY
IN <n>

Queries TTL level of pin on ‘TTL I/O’ connector. Set is not applicable.

<n> Indicates pin number on ‘TTL I/O’ connector

Valid values: 2, 3, 4, and 6.

Query response: <TTL level>

BUZZER
BUZ [ 0 | { 1 [ < n > ] } ]

Sets / queries buzzer

Set:

0 = Turn buzzer off;
1 = Turn buzzer on

if <n> specified

If <n> = 0, buzzer beeps continuously,

otherwise buzzer beeps <n> times

if <n> not specified, buzzer sounds continuously

Query response: { 0 | 1 }

0 = Buzzer off
1 = Buzzer is on continuously or beeping.

10.4.5

System Commands

SET PUMP NETWORK ADDRESS AND BAUD RATE
* ADR [ <address> [ B { 19200 | 9600 | 2400 | 1200 | 300 } ] ]
[DUAL | RECP | ALTR]

Set/query pump network address

<address> => <n> [<n>]
<address> Valid range: 0 to 99

B { 19200 | 9600 | 2400 | 1200 | 300 } will change the baud rate as indicated. NOTE: The command

response and all further communications will be at the specified baud rate.

Special communications modes

DUAL Sets the pump to Dual Pumping mode with a secondary pump.
RECP Sets the pump to Reciprocating Pumping mode with a secondary pump.
ALTR Sets Alternating pumping mode.

All special communications modes use 19,200 baud rate.

This is a special system command that will be accepted by the pump regardless of its current address or mode.
Once set, the pump will only respond to commands with the set address and at the specified baud rate.