Program function descriptions – New Era NE-1000 Multi-Phaser Programmable Syringe User Manual

Page 25

Advertising
background image

New Era Pump Systems Inc.

www.SyringePump.com

Model NE-1000 Multi-Phaser™

Publication #1200-01

20

08/01/14

9.3

Program Function Descriptions

Each program function must be in a separate Program Phase

9.3.1

‘rAtE’: Rate Function

Defines a pumping function with a fixed pumping rate. This function defines a pumping setup

consisting of the pumping rate, optional “Volume to be Dispensed”, and pumping direction. Use the ‘Rate’,
‘Volume’, and ‘Pumping Direction’ keys to set or review the pumping setup. For continuous pumping, set the
“Volume to be Dispensed” to 0.0 (off). The “Volume to be Delivered” is disabled when the display reads
[ off].

9.3.2

‘FILL’: Fill Function

The Fill function reverses the pumping direction and withdraws or dispenses the volume

dispensed or withdrawn. After a series of arbitrary dispenses, the Fill Function uses the Volume Dispensed to
automatcally refill the syringe. The Volume Dispensed is cleared when the Fill Function begins.

The pumping rate can be set, as with the Rate Function.

Fill at the current dispense/withdraw rate: If the Fill function’s pumpng rate is set to 0.0, the previous
function’s pumping rate will be used for the Fill rate.

9.3.3

‘InCr’: Increment Rate Function

The increment and decrement functions operate the same as the ‘RATE’ function, except that the

specified rate is added (‘INCR’) or subtracted (‘DECR’) from the current pumping rate. The current pumping
rate when the function is executed is the base pumping rate for the function. If no base pumping rate exists,
such as when executing a pause function or when the Pumping Program is first started, a Program error will
occur and the Program will stop.

The pumping rate units will be the same as the base pumping rate, and therefore cannot be set, nor are they
displayed, with the pumping rate increment or decrement value. As with the ‘RATE’ function, a “Volume to
be Dispensed” and pumping direction can be specified for the increment and decrement functions.

When used within a Program loop, the pumping rate can be incremented or decremented in small step
intervals.

9.3.4

‘DECr’: Decrement Rate Function

The decrement function subtracts the specified rate from the current pumping rate. For a full

description, see section 9.3.3, ‘InCr’: Increment Rate Function..

9.3.5

‘StoP’: Stop Pumping Operation and End the Program

Stops the pumping operation and stops the Pumping Program. The Pumping Program will begin

at Phase 1 when started again. An implicit ‘Stop’ function is executed when the Program exceeds the
maximum Phase number during operation.

If alarms are enabled, the buzzer will beep continuously when the Pumping Program stops.

9.3.6

‘JP:nn’: Jump to Phase

nn

The ‘Jump’ function alters the consecutive operation of Program Phases. When executed, the

Pumping Program will continue operation with Phase ‘nn’.

9.3.7

'Pr:In': Sub-Program Selection Input

The Pumping Program can be broken into sub-programs which can be selected by the user.

Auto-Run Production Mode: When Phase 1 is set to Program Selection Input, and Lockout Mode is selected,
the pump will enter Auto-Run Mode. When the pump is powered on, it will immediately begin executing the
Pumping Program. The user will immediately be prompted to enter a sub-program number. This permits a set