New Era NE-1000 Multi-Phaser Programmable Syringe User Manual

Page 35

Advertising
background image

New Era Pump Systems Inc.

www.SyringePump.com

Model NE-1000 Multi-Phaser™

Publication #1200-01

30

08/01/14

Phase

Function

Rate

Volume

Direction

11

RATE

750.0 mL/hr

10 mL

Infuse

Phase

Function

12

LP:05

Phase

Function

13

JP:01

9.4.8

Example 8: Dispensing with Complex Synchronization

:01

Repeat if
Program Input
low

:10

:10

Restart on Event Trigger
or 10 seconds

Event Trigger

W

it

h

d

ra

w

I

n

fu

s

e

TTL Program Ouput

TTL High

TTL Low

Event Trigger

Infuse until
Event Trigger

This example demonstrates a complex interaction with external equipment, such as synchronizing with another
syringe pump. The Program includes a variety of interactions with external equipment, which demonstrates
the various control possibilities of the NE-1000.

The Pumping Program begins by canceling any previous event traps (PH:01) and raising the ‘Program Output’
TTL line (PH:02). After 5.0 mL has been dispensed at 800 mL/hr (PH:03), the ‘Program Output’ TTL line is
lowered (PH:04), sending a synchronization signal to another device.

The pump then continues to pump at 800 mL/hr (PH:06) until a synchronization signal is received at the
‘Event Trigger’ TTL input, causing the Program to jump to Phase 7 (PH:05).

The pump then withdraws 0.25 mL (PH:07), pauses for 1 second (PH:08), then repeats this process if the
Program Input TTL line is low (PH:09), otherwise it continues with the next Phase.

Next, the pump pauses for 10 seconds (PH:10). Then it pauses again for the lesser of another 10 seconds
(PH:12) or until an Event Trigger occurs (PH:11). The Program then restarts (PH:13).