background image

4 How to Use

4-4. How to start the gas monitor

GX-8000

- 20 -

STEL Alarm Setpoint Display





TWA Alarm Setpoint Display





ID Display





Detection Mode





Buzzer sounds twice. (Beep, beep)






NOTE

• A sensor abnormality alarm is issued before the detection mode is entered if there is any abnormality

in the sensor. Press the RESET button. This will reset the sensor abnormality alarm temporarily, set
the gas concentration display that was abnormal on the sensor to ---, and start gas detection. However,
notify the abnormality to RIKEN KEIKI promptly. Gas for which there was an abnormality in the sensor
cannot be detected. However, the alarm cannot be reset if there is an abnormality in all the sensors.

• If there is an abnormality in the built-in clock, a fault alarm FAIL CLOCK may be issued. Press the

RESET button. The fault alarm is temporarily reset, and measurement is started with incorrect clock
time.


CAUTION

After start-up, perform air calibration before performing gas detection (air calibration mode).