4 How to Use

4-8. Display/setting mode

GX-8000

- 32 -

<Full Scale Display/Alarm Setpoint Display/Alarm Test "ALARM-P">


Display the full scale or alarm setpoint values and check the alarm operations for the settings displayed.

(1) Press the DISPLAY switch and select the "full scale display/alarm setpoint display/alarm test" from the

display/setting mode menu. The following screens are displayed in turn on the gas monitor.


(2) Press the ENTER switch to enter the alarm setpoint or other display.

NOTE

If you do not want to enter any display, press the DISPLAY switch to return to the display/setting mode
menu.

(3) Every time the ▲ or ▼ switch is pressed, the full scale and alarm setpoint menus, i.e., full scale display,

first alarm setpoint display, second alarm setpoint display, STEL alarm setpoint display, and TWA
alarm setpoint display, are displayed in turn.
Press either the ▲ or ▼ switch to select a setting that you want to check.

Select one of the following screens:

Full Scale Display

First Alarm Setpoint Display

(WARNING)

Second Alarm Setpoint Display

(ALARM)

STEL Alarm Setpoint Display (*) TWA Alarm Setpoint Display(*)

* TYPES-A, B, C, and D only

(4) Press the ENTER switch to perform alarm test. The alarm operation on this screen can be checked.

Press any switch to stop the alarm operation.


(5) Press the DISPLAY switch to exit the alarm setpoint display or alarm test. The display/setting mode

menu is displayed again.


(6) After completion, press the DISPLAY switch several times until it returns to the detection mode.