9. how to exit, Log data display "rec.data – RKI Instruments GX-8000(PT0-098) User Manual

Page 35

Advertising
background image

4 How to Use

4-9. How to exit

- 35 -

GX-8000

<Log Data Display "REC.DATA">


Display concentration data recorded to the manual memory.

(1) Press the DISPLAY switch and select the log data display from the display/setting mode menu.

The following screens are displayed in turn on the gas monitor.


(2) Press the ENTER switch to display the log data.

NOTE

If you do not want to display the log data, press the DISPLAY switch to return to the display/setting mode
menu.


(3) Every time the ▲ or ▼ switch is pressed, the log data menus are displayed in turn.

Press either the ▲ or ▼ switch to select log data that you want to check. The log data menu displays
the year, month, day, time, and memory number.


(4) Press the ENTER switch to display the selected log data.


(5) If you want to display other log data, press the ENTER switch to return to the log data menu. Repeat

the steps (3) to (5).


(6) After completion, press the DISPLAY switch to return to the detection mode.

4-9. How to exit


Make the gas monitor draw in fresh air. After the display returns to zero (or 20.9% for oxygen), keep the
POWER/ENTER switch pressed until the power is turned off.

NOTE

If the display is not zero (or 20.9% for oxygen) when the power is turned off, a
purge operation may be performed for 30 seconds at the maximum to clean the
inside of the gas monitor.