Preparing for calibration, Figure 12. calibration flow chart – RKI Instruments Beacon 200 Rig Monitor, LEL/H2S User Manual

Page 39

Advertising
background image

Beacon 200 Rig Monitor Operator’s Manual • 34

NOTE: The following procedure assumes that the target gas is present in

a high enough concentration to affect the fresh air (zero) reading.
If a fresh air environment can be verified, applying zero air to the
detector is not necessary when setting the zero reading.

This section describes how to calibrate the LEL and H

2

S detectors. It

includes procedures to prepare for calibration, set the fresh air reading, set
the span reading, and return to normal operation. It describes calibration
using a calibration kit that includes calibration gas, sample tubing, and a
fixed flow regulator with an on/off knob. RKI Instruments, Inc. recommends
using a 0.5 LPM (liters per minute) fixed flow regulator. The calibration cup/
splash guard is part of the detector assembly in the Rig Monitor.

Preparing for Calibration

1. Simultaneously press and hold the ENTER and UP/YES buttons for

approximately 5 seconds to enter the Calibration Program. Release the
buttons when the CALIBRATION PROGRAM... message displays and
asks if you want to continue or return to normal operation.

Calibrate

Channel 1 Y/N?

Calibration

Timeout Selection

ENTER

Normal Operation

Apply Gas to

Ch. 1 & Ch. 2

Detectors

Select Cal. Gas

Concentration for

Channel 2

Span

Channel 1 Y/N?

Span

Channel 2 Y/N?

Press Enter to

Adjust Span

Select Cal. Gas

Concentration for

Channel 1

Air Adjust Channel

1

Calibration

Program Enter

ENTER/ESCAPE

ESCAPE

Calibrate

Channel 2 Y/N?

Air Adjust Channel

2

Figure 12. Calibration Flow Chart