Specifications

Table 1 lists specifications for the Beacon 200 Rig Monitor.

WARNING: When using the Beacon 200 Rig Monitor, you must follow

the instructions and warnings in this manual to assure
proper and safe operation of the Beacon 200 Rig Monitor
and to minimize the risk of personal injury. Be sure to
maintain and periodically calibrate the Beacon 200 Rig
Monitor as described in this manual.

Table 1: Beacon 200 Rig Monitor Specifications

Description

Specification

Input Power

10.8 - 14.5 VDC, 1.2A VDC

Construction (housing)

Fiberglass/polyester with lexan window (NEMA 4X)

Controller Dimensions

21.0 in. H x 8.5 in. W x 7.00 in. D
(533 mm H x 216 mm W x 178 mm D)

Weight

20 lbs.

Environmental Conditions

-20°C to 50°C (-4°F to 122°F) max. ambient

Maximum humidity of 80% relative

Relays

Relay contacts rated for 10A @ 115/220V~ resistive or10A @ 30V
resistive

SPDT Form C

User Controls

Reset switch

Program buttons: ESCAPE, UP/YES, DOWN/NO, and ENTER

Detection Range

Channel 1: 0 to 100% LEL combustible gas, methane calibration standard

Channel 2: 0 to 100 PPM (parts per million) H

2

S

Response Time

90% in 30 seconds

Accuracy

LEL: ± 5% of reading or ± 2 % LEL (whichever is greater)
H

2

S: ± 5% of reading or ± 2 ppm H

2

S (whichever is greater)

Advertising