Start up – RKI Instruments 65-2399RK-CH4 User Manual

Page 12

Advertising
background image

8 • 65-2399RK-CH4 Combustible Gas Transmitter

Start Up

This section describes procedures to start up the methane transmitter and place the
transmitter into normal operation.

Introducing Incoming Power

1.

Complete the installation procedures described earlier in this manual.

2.

Verify that the power wiring to the controller is correct and secure. Refer to the
controller operator’s manual.

3.

Turn on power to the controller.

4.

Turn on the controller.

5.

Verify that the controller is on and operating properly. Refer to the controller
operator’s manual.

NOTE:

When first powered up, the transmitter will enter about a one minute period
when the 4-20 mA output is stabilizing and may be above the controller alarm
points or well below zero momentarily. RKI controllers have a one minute
warmup period when the controller does not display any gas reading or give any
alarm indication. The methane transmitter’s 4-20 mA signal should be stable by
the time the controller’s warmup period is over.

CAUTION:

Allow the detector to warm up for 5 minutes before you continue with the next
section, “Setting the Zero Signal”.

Setting the Zero Signal

NOTE:

If you can verify that the methane transmitter is in a fresh air environment, you
do not need to apply zero air to the detector before adjusting the zero reading.

The procedure below describes applying zero emission air, usually called zero air, using a
calibration kit that includes a calibration cup, calibration gas, sample tubing, and a fixed
flow regulator with an on/off knob. RKI Instruments, Inc. recommends using a 0.5 LPM
(liters per minute) fixed flow regulator.

1.

Unscrew and remove the junction box cover from the junction box.

2.

Set a voltmeter to measure in the millivolt (mV) range.

3.

Plug the voltmeter leads into the test points on the amplifier. Plug the positive lead
into the red (+) test point; plug the negative lead into the black (-) test point.

4.

Screw the calibration cup onto the bottom of the IR CH

4

%Vol. detector.

5.

Screw the regulator into the zero air calibration cylinder.

6.

Use the calibration kit sample tubing to connect the regulator to the calibration cup.

7.

Turn the regulator’s on/off knob counterclockwise to open it. Gas will begin to flow.

8.

Allow the gas to flow for one minute.

9.

Verify a voltmeter reading of 100 mV (±2 mV).

10. If necessary, use a small flat-blade screwdriver to adjust the zero pot until the

Advertising