RKI Instruments 65-2399RK-CH4 User Manual

Page 20

Advertising
background image

16 • 65-2399RK-CH4 Combustible Gas Transmitter

NOTE:

The following procedure assumes the use of a calibration kit which includes a
calibration gas cylinder, a 0.5 LPM fixed flow regulator with an on/off knob, a
calibration cup for the detector, and a short piece of sample tubing to connect the
regulator to the calibration cup.

Preparing for Calibration

1.

Unscrew and remove the junction box cover.

2.

Set a voltmeter to measure in the millivolt (mV) range.

3.

Plug the voltmeter leads into the test points on the amplifier. Plug the positive lead
into the red (+) test point; plug the negative lead into the black (-) test point.

4.

Use the following formula to determine the correct test points output for the
calibrating sample.

Output (mV) = (calibrating sample/fullscale) X 400 + 100

For example, with a calibrating sample of 50 %Vol. and a fullscale setting of
100 %Vol., the correct output is 300 mV.

300(mV) = (50/100) X 400 +100

5.

Screw the calibration cup onto the detector housing.

6.

Place the controller into its calibration mode or disable external alarms.

NOTE:

Calibrating the methane transmitter may cause alarms. Be sure to put the
controller into its calibration mode or disable external alarms before continuing.

Setting the Zero Reading

NOTE:

If you can verify that the methane transmitter is in a fresh air environment, you
do not need to apply zero air to the detector before adjusting the zero reading.

1.

Screw the regulator into the zero air calibration cylinder.

2.

Use the calibration kit sample tubing to connect the regulator to the calibration cup.

3.

Turn the regulator’s on/off knob counterclockwise to open the regulator.

4.

Allow the gas to flow for one minute and verify a reading of 100 mV (±2 mV). If
necessary, use the zero pot on the amplifier to adjust the reading to 100 mV (±2 mV).

5.

Turn the regulator’s on/off knob clockwise to close the regulator.

6.

Unscrew the regulator from the zero air calibration cylinder.

7.

Leave the sample tubing connected to the regulator and the calibration cup.

Setting the Response Reading

1.

Screw the regulator into the calibration cylinder. Verify that the calibration gas is
representative of the transmitter’s target gas.

2.

Turn the regulator’s on/off knob counterclockwise to open the regulator.

3.

Allow the calibration gas to flow for one minute and verify that the reading matches
the response reading (±2 mV) you determined earlier. If necessary, use the span pot on
the amplifier to adjust the reading to match the correct response reading.

Advertising