Troubleshooting – RKI Instruments 65-2399RK-CH4 User Manual

Page 15

Advertising
background image

65-2399RK-CH4 Combustible Gas Transmitter • 11

Troubleshooting

The troubleshooting guide describes symptoms, probable causes, and recommended
action for problems you may encounter with the methane transmitter.

NOTE:

This troubleshooting guide describes transmitter problems only. See the
controller operator’s manual for problems you may encounter with the
controller.

Table 2:Troubleshooting the Methane Transmitter

Condition

Symptom(s)

Probable Causes

Recommended Action

Fail Condition

• Controller indicates a

fail condition.

• The transmitter wiring

is disconnected or
misconnected.

• The wiring from the

detector to the
amplifier is
disconnected or
misconnected.

• The plug-in sensor is

not properly plugged
into the 3 socket
pattern in the detector
housing body.

• The transmitter’s zero

reading is low enough
to cause a fail
condition.

• The transmitter is

malfunctioning.

1. Verify that the transmitter wiring is

correct and secure.

2. Verify that the wiring from the detector

to the amplifier is correct and secure.

3. Confirm that the plug-in sensor is

properly installed.

4. Perform a zero adjustment. A full

calibration is recommended.

5. If the fail condition continues, replace

the sensor as described later in this
section.

6. If the fail condition continues, contact

RKI for further instruction.

Slow or No
Response/
Difficult or
Unable to
Calibrate

• Transmitter responds

slowly or does not
respond to response
test.

• Unable to accurately

set the zero or
response reading
during calibration.

• Transmitter requires

frequent calibration.

Note: Under “normal”
circumstances, the
transmitter requires
calibration once every 6
months.

Some applications

may require a more
frequent calibration
schedule.

• The calibration cylinder

is low, out-dated, or
defective.

• The calibration gas is

not an appropriate
concentration.

• The membrane on the

detector housing cap is
blocked with dirt or
some other particulate
matter.

• The transmitter is

malfunctioning.

1. Verify that the calibration cylinder

contains an adequate supply of a
fresh test sample.

2. Verify that the calibration gas

concentration is appropriate for the
transmitter.

3. Check the face of the detector

housing cap and remove any
particulate contamination from the
hydrophobic membrane if necessary.
If membrane appears saturated with
contamination or damaged, replace
the membrane as described in the
next section.

4. If the calibration/response difficulties

continue, replace the sensor as
described later in this section.

5. If the calibration/response difficulties

continue, contact RKI for further
instruction.

Advertising