Maintenance – RKI Instruments 65-2399RK-CH4 User Manual

Page 13

Advertising
background image

65-2399RK-CH4 Combustible Gas Transmitter • 9

voltmeter reading is 100 mV (±2 mV).

11. Turn the regulator’s on/off knob clockwise to close it.

12. Unscrew the calibration cup from the detector.

13. Unscrew the regulator from the zero air calibration cylinder. For convenience, leave

the sample tubing connected to the regulator and the calibration cup.

14. Store the components of the calibration kit in a safe and convenient place.

15. Remove the voltmeter leads from the test points.

16. Secure the junction box cover to the junction box.

Maintenance

This section describes maintenance procedures. It includes preventive maintenance,
troubleshooting, and component replacement procedures.

Preventive Maintenance

This section describes a preventive maintenance schedule to ensure the optimum
performance of the methane transmitter. It includes daily, monthly, and biannual
procedures.

Daily

Verify a display reading of 0 %Vol. at the controller. Investigate significant changes in the
display reading.

Monthly

This procedure describes a test to verify that the methane transmitter responds properly
to the target gas.

WARNING: The controller is not an active gas monitoring device during the response

test procedure.

NOTE:

Performing a response test on the methane transmitter may cause alarms. Be sure
to put the controller into its calibration mode or disable external alarms before
performing this test.

NOTE:

The following procedure assumes the use of a calibration kit which includes a
calibration gas cylinder, a 0.5 LPM fixed flow regulator with an on/off knob, a
calibration cup for the detector, and a length of sample tubing.

Preparing for the response test

1.

Place the controller into its calibration mode or disable external alarms.

2.

Verify that the controller display reading for the channel you are testing is 0 %Vol.

If the display reading is not 0 %Vol., set the zero reading of the transmitter as
described in “Start Up” on page 8, then continue this procedure.

3.

Screw the regulator into the calibration cylinder.

4.

Screw the calibration cup onto the bottom of the detector.

Advertising