[ CARE AND USE MANUAL ]

iKey

■

■

Don’t assume the iKey is to blame:

■

■

Investigate cause of iKey failure;

■

■

Monitor backpressure;

■

■

Examine mobile phase age, bacterial contamination, mobile phase precipitation, etc.;

■

■

Examine sample quality;

■

■

Examine injection solvent strength.

■

■

Do not prepare excessive amounts of mobile phase:

■

■

To reduce the chances of mobile phase contamination or degradation, prepare enough mobile phase to last for 3–4 days.
Alternatively, store excess bulk quantities in a refrigerated environment.

b. Troubleshooting Questions

■

■

What is the age of your 100% aqueous mobile phase?

■

■

Is the mobile phase filtered through a 0.2 µm membrane?

■

■

Was the mobile phase prepared fresh or topped off?

■

■

Is the water source of adequate quality?

■

■

When was the last time the water system was serviced or was the bottle of water unopened?

■

■

Is bacterial growth a possibility?

■

■

If a neat standard is prepared in the initial mobile phase conditions and injected, are the problems still observed?

■

■

If the sample is additionally filtered/purified (i.e., SPE, filtration, etc.) is the problem still observed?

■

■

Has the quality of the samples changed over time?

c. Good Practices

■

■

Do not operate at pressures that exceed 10000 psi (690 bar).

■

■

Do not apply electrospray potential to the emitter for an extended period of time without mobile phase flow.

■

■

Do not immerse in liquid.

■

■

Do not touch any part of the iKey other than the handle.

■

■

Do not freeze.

■

■

Flush with acetonitrile and store at room temperature when not in use.

■

■

Use the iKey sheath to protect device when not in use.

■

■

Do not bend or pull the capillary connection tubing at the ionKey source coupling.

■

■

Avoid excess voltage (>4.0 kV) which can erode emitter over time.