Waters 30 mm ID Preparative Guard Cartridges and Guard Holders User Manual

Care and use manual

Advertising
background image

[ CARE AND USE MANUAL ]

30 mm ID Preparative Guard Cartridges and Guard Holders

CONT ENTS

I.

INTRODUCTION

II.

SPECIFICATIONS

III.

TIGHTENING PROCEDURE

IV.

START-UP RECOMMENDATIONS

a. Background (Avoid Pressure Pikes)
b. Start-Up

V.

INSTALLING AND REPLACING
THE GUARD CARTRIDGE AND
CARTRIDGE HOLDER

VI.

WHEN TO REPLACE THE
GUARD CARTRIDGE

VII.

REPLACING THE O-RINGS

VIII. TROUBLESHOOTING

IX.

GUARD CARTRIDGE STORAGE

a. HPLC Guard Cartridges
b. SFC Guard Cartridges

X.

ORDERING INFORMATION

XI.

WARRANT Y/SERVICE INFORMATION

I. INTRODUCTION

Thank you for purchasing the 30 mm ID Preparative Guard Cartridge and
Guard Holder assembly. This preparative guard assembly is designed to
protect 30 mm ID Preparative Optimum Bed Density (OBD™) Columns and
is suitable for both HPLC and SFC.

II. SPECIFICATIONS

Preparative Guard Cartridge Specifications

Dimension

30 mm ID x 10 mm length

Pressure Limit

6,000 psi (413 bar)

Materials Construction

316 stainless steel

Solvent Compatibility

Suitable for common HPLC (reversed-phase and
normal-phase) and SFC mobile phases. Minor
swelling of O-rings in THF and chloroform may
require more frequent replacement if leaks
observed. Solvent compatibility of the packing
is the same as that of the preparative column.

Column Volume

7 mL

Advertising