IX. GUARD CARTIDGE STORAGE

a. HPLC Guard Cartridges

To maintain the integrity of the bonded phase as well as the guard cartridge hardware and frits, flush
the guard cartridge with at least eight column volumes of 100% acetonitrile after every use. Never
store the guard cartridge in methanol or aqueous methanol. If the mobile phase contains a buffer salt,
first flush the guard cartridge with at least ten column volumes of 18 MegaOhm water followed by at
least eight column volumes of 100% acetonitrile. Please refer to Section II for the column volume.

b. SFC Guard Cartridges

Purging the SFC prep guard cartridge in pure CO

2

is recommended prior to storage. This will eliminate

or minimize retention changes of the guard cartridge during storage. Storing the guard cartridge in
methanol is not recommended.

VIII. TROUBLESHOOTING

Troubleshooting Tips

Symptom

Cause

Corrective Action

Solvent leak at seam
of cartridge weep hole

Cartridge not seated in end fitting

Disassemble, properly seat cartridge

Particles on the face of the
O-ring or on metal seat

Wipe clean with a cotton tip applicator

Worn O-ring

Replace O-ring

Broad peaks, loss
of peak resolution

Packing material fouled or voided

Replace with new guard cartridge

Guard cartridge packing material
different from prep column packing

Replace with matching guard cartridge

High backpressure

Filters in guard cartridge
plugged or fouled

Replace with new guard cartridge

Guard cartridge chemically fouled

Replace with new guard cartridge