c. Column Use Information

The eCord Chip provides the customer with column use data, column dimensions and serial number. The overall
column usage information includes the total number of samples, total number of injections, total sample sets,
date of first injection, date of last injection, maximum pressure and temperature. The information also details
the column history by sample set including date started, sample set name, user name, system name, number
of injections in the sample set, number of samples in the sample set, maximum pressure and temperature in
the sample set, and if the column met basic system suitability requirements. Up to 50 sample sets can be
stored on the eCord Chip. In addition, the eCord provides two-way communications between the eCord Chip and
Empower

®

Software.