Care and use manual – Waters XSelect CSH130 C18 3.5 μm and 5 μm Columns User Manual
Page 10

[ CARE AND USE MANUAL ]
XSelect CSH130 C
18
Columns
10
1.20
1.10
1.00
0.90
0.80
0.70
0.60
0.50
0.40
0.30
0.20
0.10
0.00
2
3
4
5
6
7
8
9
10
11
12 Min.
Figure 10: Representative Test Chromatography and Conditions for Separation of a Cytochrome c Tryptic Digest
Peak Identification
T1
N-AcGDVEK
T13-T14
KYIPGTK
T14
YIPGTK
T4
IFVQK
T9-T10
KTGQAPGFSYTDANK
T10
TGQAPGFSYTDANK
T8
TGPNLHGLFGR
T15
MIFAGIK
T19C
EDLIAY
T19
EDLIAYLK
T12-T13
GITWGEETLMEYLENPKK
T12
GITWGEETLMEYLENPK
T5
CAQCHTVEK (heme attached)
Retention Time (min)
Result
T19C
6.69
Retention Time Difference (min)
Result
T4 and T14
0.312
T10 and T9-T10
0.780
T19 and T19C
0.461
T12 and T12-T13
0.290
Chromatographic Conditions
Column:
2.1 mm x 50 mm
Sample:
Tryptic digest of bovine cytochrome c (P/N 186006371)
Reconstituted w/ 200µL of MP A, 5.0µL injection
Flow rate:
0.2 mL/min
Temperature: 40°C
Mobile Phase A: 0.1% formic acid in 100% water MPA
Mobile Phase B: 0.085% formic acid in 75% acetonitrile MPB
Gradient:
1 to 21%B in 5 min; 21 to 31.4% B in 2.5 min; 31.4 to
95% B in 2.5 min; 95% B for 1.5min; 1%B for 2.5 min
Detection:
UV 214 nm