a. Tips for maximizing ACQUITY UPLC CSH Column
Lifetimes

1. To maximize ACQUITY UPLC CSH column lifetime, pay close

attention to:

Water quality (including water purification system)

Solvent quality

Mobile-phase preparation, storage and age

Sample, buffer and mobile-phase solubilities

Sample quality and preparation.

2. When problems arise, often only one improper practice must

be changed.

3. Always remember to:

Use in-line filter unit or, preferably, a VanGuard Pre-column.

Discourage bacterial growth by minimizing the use of 100%

aqueous mobile phases where possible.

Change aqueous mobile phase every 24-48 hours (if 100%

aqueous mobile phase use is required).

•

Discard old 100% aqueous mobile phases every 24-48 hours to

discourage bacterial growth.

•

Add 5%-10% organic modifier to mobile phase A and adjust

gradient profile.

•

Filter aqueous portions of mobile phase through 0.2 µm filter.

•

Maintain your water purification system so that it is in good

working order.

•

Only use ultra pure water (18 megohm-cm) water and highest

quality solvents possible. HPLC grade water is not UPLC grade

water.

•

Consider sample preparation (e.g., solid-phase extraction,

filtration, etc).

4. Avoid (where possible):

100% aqueous mobile phases (if possible).

HPLC-grade bottled water.

“Topping off” or adding “new” mobile phase to “old” mobile phase.

Old aqueous mobile phases. Remember to rinse bottles

thoroughly and prepare fresh every 24-48 hours.

Using phosphate salt buffer in combination with high ACN

concentrations (e.g., > 70%) due to precipitation.

5. Don’t: assume a “bad” column is the culprit when high backpressure or

split peaks are observed.

Investigate cause of column failure:

Backpressure

Mobile phase(s), bacteria, precipitation and/or samples

Peak splitting

Sample quality

Injection solvent strength.

6. Remember: UPLC flow rates are often much lower and, therefore,

mobile phases last much longer (only prepare what you need or store
excess refrigerated).

7. Mobile-phase related questions to ask:

Am I using 100% aqueous mobile phases? Am I able to add a

small amount of organic modifier to my mobile-phase A?

Do I filter my aqueous mobile phases through 0.2 µm filters?

How old is my mobile phase? Do I label the bottle with

preparation date?

Do I “top off” or do I prepare fresh mobile phases every 24-48

hours?

What is the quality of my water? Has the quality recently

changed? How is my water purification system working? When

was it last serviced?

Am I working with a pH 7 phosphate buffer (which is VERY

susceptible to bacterial growth)?

8. Sample-related questions to ask:

If I inject neat standards prepared in mobile phase do I observe

these problems?

If I prepare my standards in water and prepare them like samples

(e.g., SPE, filtration, etc.) do I still observe these problems?

Has the quality of my samples changed over time?