IV. USING THE Hi3 DIGESTION STANDARDS 
 
FOR QUANTIFICATION
For a typical proteomic analysis carried out with 75 µ i.d. UPLC
®
Columns, sufficient Hi3 standard should be added to each sample to 
provide ca. 50 fmol of standard per injection. The injected quantity 
should be scaled for other column diameters. LC/MS
E
data should be
obtained in the normal manner.
For data processing, first append the appropriate protein sequence 
below to the databank to be searched for protein identification 
using a text editor.
>CLPB_ECOLI (P03815) ClpB protein (Heat shock protein F84.1)
MRLDRLTNKFQLALADAQSLALGHDNQFIEPLHLMSALLNQEGGSVS 
PLLTSAGINAGQLRTDINQALNRLPQVEGTGGDVQPSQDLVRVLNLCDK 
LAQKRGDNFISSELFVLAALESRGTLADILKAAGATTANITQAIEQMRG 
GESVNDQGAEDQRQALKKYTIDLTERAEQGKLDPVIGRDEEIRRTIQVL 
QRRTKNNPVLIGEPGVGKTAIVEGLAQRIINGEVPEGLKGRRVLALDMG 
ALVAGAKYRGEFEERLKGVLNDLAKQEGNVILFIDELHTMVGAGKADG 
AMDAGNMLKPALARGELHCVGATTLDEYRQYIEKDAALERRFQKVFV 
AEPSVEDTIAILRGLKERYELHHHVQITDPAIVAAATLSHRYIADRQLP 
DKAIDLIDEAASSIRMQIDSKPEELDRLDRRIIQLKLEQQALMKESDEAS 
KKRLDMLNEELSDKERQYSELEEEWKAEKASLSGTQTIKAELEQAKIAI 
EQARRVGDLARMSELQYGKIPELEKQLEAATQLEGKTMRLLRNKVTDA 
EIAEVLARWTGIPVSRMMESEREKLLRMEQELHHRVIGQNEAVDAVSN 
AIRRSRAGLADPNRPIGSFLFLGPTGVGKTELCKALANFMFDSDEAMV 
RIDMSEFMEKHSVSRLVGAPPGYVGYEEGGYLTEAVRRRPYSVILLDEV 
EKAHPDVFNILLQVLDDGRLTDGQGRTVDFRNTVVIMTSNLGSDLIQER 
FGELDYAHMKELVLGVVSHNFRPEFINRIDEVVVFHPLGEQHIASIAQIQ 
LKRLYKRLEERGYEIHISDEALKLLSENGYDPVYGARPLKRAIQQQIENP 
LAQQILSGELVPGKVIRLEVNEDRIVAVQ 
>PHS2_RABIT (P00489) GLYCOGEN PHOSPHORYLASE, MUSCLE 
FORM (ECSRPLSDQEKRKQISVRGLAGVENVTELKKNFNRHLHFTLVK 
DRNVATPRDYYFALAHTVRDHLVGRWIRTQQHYYEKDPKRIYYLSLE 
FYMGRTLQNTMVNLALENACDEATYQLGLDMEELEEIEEDAGLGNGGL 
GRLAACFLDSMATLGLAAYGYGIRYEFGIFNQKICGGWQMEEADDWL 
RYGNPWEKARPEFTLPVHFYGRVEHTSQGAKWVDTQVVLAMPYDTP 
VPGYRNNVVNTMRLWSAKAPNDFNLKDFNVGGYIQAVLDRNLAENIS 
RVLYPNDNFFEGKELRLKQEYFVVAATLQDIIRRFKSSKFGCRDPVRTN 
FDAFPDKVAIQLNDTHPSLAIPELMRVLVDLERLDWDKAWEVTVKTCA 
YTNHTVLPEALERWPVHLLETLLPRHLQIIYEINQRFLNRVAAAFPGDV 
DRLRRMSLVEEGAVKRINMAHLCIAGSHAVNGVARIHSEILKKTIFKDFY 
ELEPHKFQNKTNGITPRRWLVLCNPGLAEIIAERIGEEYISDLDQLRKLLS 
YVDDEAFIRDVAKVKQENKLKFAAYLEREYKVHINPNSLFDVQVKRIH 
EYKRQLLNCLHVITLYNRIKKEPNKFVVPRTVMIGGKAAPGYHMAKMII 
KLITAIGDVVNHDPVVGDRLRVIFLENYRVSLAEKVIPAADLSEQISTAG 
TEASGTGNMKFMLNGALTIGTMDGANVEMAEEAGEENFFIFGMRVEDV 
DRLDQRGYNAQEYYDRIPELRQIIEQLSSGFFSPKQPDLFKDIVNMLMH 
HDRFKVFADYEEYVKCQERVSALYKNPREWTRMVIRNIATSGKFSSDR 
TIAQYAREIWGVEPSRQRLPAPDEKIP 
Configure the modified databank with the PLGS Databank Admin Tool.
Create a search workflow using the PLGS Workflow Designer, 
being sure that the modified databank is selected, the proper 
identifier for the standard protein is entered on the Calibration 
Protein line and the number of femtomoles injected is entered 
on the Calibration Protein Concentration line, as below: