Waters XBridge Direct Connect HP 2.1 x 30 mm Column for Automated Sample Preparation with Waters UPLC with Online SPE Technology User Manual
Care and use manual, Ab c
 
[ Care and Use ManUal ]
XBridge Direct Connect HP 2.1 x 30 mm Column
1
The XBridge
™
Direct Connect HP 2.1 x 30 mm column makes it possible
to analyze specific analytes in a sample matrix when combined with 
Waters ACQUITY UPLC
®
columns and online SPE technology.
CONTENTS
I.
INTRODUCTION
II. SAMPLE PREPARATION
III. EQUIPMENT
IV. TYPICAL CHROMATOGRAPHIC CONDITIONS
1. INTRODUCTION
The XBridge packing materials were designed to provide excellent 
peak shape, high efficiency, and excellent stability for acidic and basic 
mobile phases. The recommended operating pH limits for XBridge Direct 
Connect HP columns are pH 1-12. For more information on XBridge 
HPLC analytical columns, please refer to the XBridge Columns Care and 
Use Manual (P/N 720001366EN). The XBridge Direct Connect column 
can be screwed directly into the switching valve or connected to fit-
tings like a conventional HPLC column.
XBRIDgE DIRECT CONNECT HP 2.1 x 30 mm COlUmN fOR AUTOmATED SAmPlE 
PREPARATION wITH wATERS UPlC wITH ONlINE SPE TECHNOlOgy
II. SAmPlE PREPARATION
Aqueous Samples: Water, Beverages
Adjust pH to maximize analyte retention on the SPE sorbent. Buffer 
salts and dispersive agents may be used to increase partitioning onto 
the SPE sorbent. Pretreatment to remove suspended matter prior to 
SPE treatment may include filtration or centrifugation.
Non-Aqueous Liquid
When appropriate the sample may be diluted with aqueous buffers 
and organic co-solvents for reversed-phased SPE. If sufficient dilution 
has occurred, the sample may be treated in a manner similar to an 
aqueous sample.
III. EQUIPmENT
Shown in Figure 1 is a schematic of Waters UPLC
®
system with online
SPE technology, which consists of a 2777 high-pressure sample 
manager, an ACQUITY UPLC
®
quaternary solvent manager (QSM), an
ACQUITY UPLC binary solvent manager (BSM), two extraction columns, 
an analytical column, and detection system (mass spectrometry, PDA 
or fluorescence). The extraction columns are connected to a switching 
valve so they can be operated in parallel.
2777c
Detector
ACQUITY
QSM
ACQUITY
BSM
SPE
Injection
LC
Injection
Extraction
Column A
ACQUITY BEH Column
Extraction
Column B
Waste
2.0 mL/min
0.4 mL/min
Valve A: Injection valve for online SPE LC-MS/MS
Valve B: switching valve for extraction columns
Valve C: injection valve for stand-alone LC-MS/MS
A
B
C
Figure 1: Schematic of Waters UPLC System with Online SPE Technology.