Care and use manual – Waters Protein Separation Technology ACQUITY UPLC BEH300, C4, 1.7 µm Columns User Manual
Page 9
 
[ Care and Use ManUal ]
Protein Separation Technology ACQUITY UPLC BEH300, C
4
, 1.7
�m
9
b. Storage
For periods longer than four days at room temperature, store the 
column in 100% acetonitrile. Immediately after use with elevated 
temperatures and/or at pH extremes, store in 100% acetonitrile for 
the best column lifetime. Do not store columns in highly aqueous 
(<20% organic) mobile phases, as this may promote bacterial growth. 
If the mobile phase contained a buffer salt, flush the column with 
10 column volumes of HPLC grade water (see Table 1 for common 
column volumes) and replace with 100% acetonitrile for storage. 
Failure to perform this intermediate step could result in precipitation 
of the buffer salt in the column or system when 100% acetonitrile is 
introduced. Completely seal column to avoid evaporation and drying 
out of the bed. 
Note: If a column has been run with a mobile phase that contains 
formate (e.g., ammonium formate, formic acid, etc.) and is then 
flushed with 100% acetonitrile, slightly longer equilibration times 
may be necessary when the column is re-installed and run again with 
a formate-containing mobile phase.
VII. IntroduCInG eCord InteLLIGent CHIP teCHnoLoGY
a. Introduction
The eCord intelligent chip is a new technology that will provide 
the history of a column’s performance throughout its lifetime. The 
eCord will be permanently attached to the column to assure that the 
column’s performance history is maintained in the event that the 
column is moved from one instrument to another.
Figure 3: Waters eCord Intelligent Chip
At the time of manufacture, tracking and quality control information 
will be downloaded to the eCord. Storing this information on the chip 
will eliminate the need for a paper Certificate of Analysis. Once the 
user installs the column, the software will automatically download 
key parameters into a column history file stored on the chip. In 
this manual, we explain how the eCord will provide a solution for 
easily tracking the history of the columns, reduce the frustration of 
paperwork trails, and give customers the reassurance that a well 
performing column is installed onto their instruments.
Figure 4: eCord Inserted into Side of Column Heater
b. Installation
Install the column into the column heater. Plug the eCord into the 
side of the column heater. Once the eCord is inserted into the column 
heater the identification and overall column usage information will 
be available in ACQUITY Console, allowing the user to access column 
information on their desktop.
c. Manufacturing Information