Care and use manual – Waters Certified Sep-Pak User Manual

Page 4

Advertising
background image

[ Care and Use ManUal ]

Certified Sep-Pak Cartridges

4

5. Elute

Once the interferences are washed off the cartridge, a strong solvent is introduced to elute the analytes of interest. The volume and flow rate
of the eluting solvents should be precisely controlled as in the load step to ensure reproducible results.

Reversed Phase

Normal Phase

Ion Exchange

Analyte

Low to moderate polarity/

hydrophobic

Moderate to highly polar/uncharged

Charged or ionized

Matrix

Aqueous

Non-polar organic solvent

Aqueous/low ionic strength

Condition/Equilibrate

1. Solvate polar organic

2. Water

Non-polar organic

Low ionic strength buffer

Wash

Aqueous/buffer

Non-polar

Low ionic strength buffer

Elute

Increase polar organic content

in steps

Increase moderate to high polarity

organic content in steps

Stronger buffers ionic strength or

pH to neutralize the charge

Table 2. Guidelines on the Various Types of Separation Mechanisms

IV. FACtoRs AFFeCtInG sePARAtIon eFFICIenCY

Separation efficiency varies with flow rate and sample load. Here
are some guidelines to consider when optimizing your separation.

a. Flow Rate

If the sample flow rate is too high, components may not interact
sufficiently with he sorbent. The result is loss of resolution, analyte
breakthrough, and poor recovery. To condition the cartridge, use a
flow rate of 5-10 mL/min. To load and elute Certified Sep-Pak Vac 3
cc 200 mg, and 500 mg cartridges, use a flow rate or 0.2-1 mL/min.
To load and elute all other cartridges, use a flow rate or 2-10 mL/min.

b. Sample Load

Cartridge capacity depends upon the mass of all components in the
sample, volume of solvent, solvent strength, sorbent type, sorbent
quality, and retention factor (k) of components. Overloading results in
sample breakthrough which leads to variability in sample recovery.

V. stoRAGe AnD DIsPosAL oF UseD CARt RIDGes

Cartridges stored in their original sealed pouch remain stable
for long periods. Once opened, special care must be taken to
ensure that the SPE cartridges are not exposed to the laboratory
environment for long periods. The highly active sorbents will readily
accumulate background contaminants from the environment and
will affect the quality of the device. Waters recommends that you
open and use only those cartridges that will be needed. If this is not
possible, store the exposed cartridges in clean contaminant free
container such as a desiccator.

Dispose of used cartridges safely in accordance with applicable
government or local regulations.

Advertising