Care and use manual – Waters ACQUITY UPLC BEH Glycan, 1.7 µm Columns, Glycan Performance Test Standard User Manual
Page 3
 
[ CARE AND USE MANUAL ]
ACQUITY UPLC BEH Glycan, 1.7 �m Columns
3
4. Gradually increase the flow rate from 0.25 to 0.5 mL/min
over 3 minutes.
5. Once a stable backpressure and baseline have been achieved,
proceed to the next section.
d. Column Equilibration
Glycan Separation Technology columns are shipped in 100% 
acetonitrile. It is important to ensure mobile phase compatibility 
before changing to a different mobile-phase system. Equilibrate 
the column with a minimum of 10 column volumes of the mobile 
phase to be used (refer to Table 1 for column volumes). 
Table 1: Empty Column Volumes in mL 
(multiply by 10 for flush solvent volumes) 
Column Length
Volume in mL (2.1 mm i.d.)
50
0.17
100
0.35
150
0.52
 
To avoid precipitating mobile-phase buffers on your column or in your 
system, flush the column with five column volumes of a water/organic 
solvent mixture using the same or higher acetonitrile content as in 
the desired buffered mobile phase. For example, flush the column and 
UPLC system with 50% acetonitrile in water prior to introducing 50% 
acetonitrile/50% buffered mobile phase.
Column equilibration may be judged initially by stable pressure and 
by a stable detector baseline. For a specific application, it is, however, 
necessary to test the required duration of equilibration. The criteria 
for adequate equilibration include reproducibility of retention time 
for major and minor peaks, resolution for critical pairs, and consistent 
baseline characteristics.
Note: Low concentration mobile-phase additives, particularly those 
with minimal buffering capacity, may require extended equilibration 
and re-equilibration between gradient analyses.
e. Initial Column Efficiency Determination
1. Perform an efficiency test on the column before using it in the
desired application. Waters recommends using the solute mixture 
and conditions described in the “Performance Test Chromatogram” 
to test the column upon receipt. These conditions can be found on 
the eCord attached to the column.
2. Measure the retention of the test compounds and the number of
theoretical plates (N).
3. Repeat the test at predetermined intervals to track column
performance over time. Slight variations may be obtained on 
two different UPLC systems due to the quality of the connections, 
operating environment, system electronics, reagent quality, 
condition of column, and operator technique.
f. Useful Functional Tests for Benchmarking a New Column
The performance test uses a Glycan Performance Test Standard 
(Part No. 186006349). This test sample consists of 2-AB Human 
IgG and is QC verified to contain the components needed to 
benchmark and monitor a new column, as shown in Figure 1. 
To prepare the standard, add 100 µL of 100mM ammonium formate 
buffer pH 4.5 and 100 µL of acetonitrile directly to the vial for a 
total volume of 200 µL.
Gently mix the sample by inversion.
Human IgG Glycans Performance Test
The conditions shown below are for use on an ACQUITY UPLC system. 
If you are using an ACQUITY UPLC H-Class or ACQUITY UPLC H-Class 
Bio system, use 50/50 acetonitrile/water for the purge and wash 
solvents. All of the other conditions can remain the same. Note that the 
ACQUITY UPLC H-Class and ACQUITY UPLC H-Class Bio systems do 
not have a strong and weak needle wash. Instead, they have one 
purge and one wash solvent, both of which should be in 50/50 
acetonitrile/water. 
It is also important to note the injection solvent and injection volume 
for this application. Larger glycans have limited solubility in solutions 
that contain more than 50% acetonitrile. Gradual precipitation and 
loss of these larger glycans will be observed under these condtions. 
However, also note that large injections of water-containing samples 
will distort the peak shape in HILIC chromatography. The optimum 
injection volume for these applications is < 3 µL.