Care and use manual – Waters µStyragel Column User Manual
Page 2
 
[ Care and Use ManUal ]
µStyragel Column
2
Band Spreading
There is a small degree of band spreading due to mechanical causes 
inherent in any chromatographic system. Connections, tubing, and 
equipment performance all contribute to instrument band spreading. 
It is important to know the degree of instrument band spreading in 
your system.
Before installing the column, measure your system’s instrument band 
spreading (see Section V., Testing Band Spreading). If this test is not 
possible with your system, refer to your system operator’s manual.
b. Connecting the Column to the GPC System
Handle the column with care. Do not drop or hit the column on a hard 
surface as it may disturb the bed and affect its performance.
1. Correct connection of 1/16-inch outer diameter stainless steel
tubing leading to and from the column is essential for high-
quality chromatographic results.
2. When using standard stainless steel compression screw fittings,
it is important to ensure proper fit of the 1/16-inch outer diam-
eter stainless steel tubing. When tightening or loosening the 
compression screw, place a 5/16-inch wrench on the compres-
sion screw and a 3/8-inch wrench on the hex head of the column 
endfitting.
Note: If one of the wrenches is placed on the column flat during this 
process, the endfitting will be loosened and leak.
3. If a leak occurs between the stainless steel compression screw
fitting and the column endfitting, a new compression screw 
fitting, tubing and ferrule must be assembled.
4. An arrow on the column identification label indicates correct
direction of solvent flow.
Column Connectors
Due to the absence of an industry standard, various column manu-
facturers have employed different styles of chromatographic column 
connectors. The chromatographic performance of your separation can 
be negatively affected if the style of your column endfittings do not 
match the existing instrumentation tubing ferrule setting.
The two styles of endfittings differ by the length of tubing that 
protrudes from the ferrule. Parker style endfittings require a 
0.090-inchferrule depth. Waters style endfittings (e.g., µStyragel) 
require a 0.130-inch ferrule depth. Examples of each endfitting are 
given below.
Figure 1: Waters Ferrule Setting
Figure 2: Parker Ferulle Settings
In a proper tubing/column connection the tubing touches the bottom 
of the column endfitting with no void between them. 
Figure 3: Proper Tubing/Column Connection
c. Connecting Columns in Series
When connecting columns in series, use the 0.009-inch (0.25 
mm) i.d. U-shaped column joining tube supplied with each column. 
Guidelines for columns with different pore sizes when using columns 
of different pore sizes in series: 
• Place the 100Å column last in the series to minimize pres-
sure pulses likely to result from highly viscous samples.
• Place the 10
6
Å column as the last column in a series,
except	when	using	it	in	conjunction	with	a	100Å	column.	
If	you	use	these	columns	together,	place	the	100Å	column	
last	and	the	100Å	column	next	to	last.	For	analytical	
separations, the remaining columns can be put in any order.
• When operating in the preparative mode, connect columns
in	order	of	decreasing	pore	size	(105Е,	104Е,	100Е,	
500Е)	to	achieve	maximum	loading.
0.130 inches
0.090 inches