Care and use manual, General considerations, B. efficiency testing – Waters IC-Pak C Method Development Column User Manual
Page 4: Preparing mobile phase, Guidelines for the calibration standard, Preparing individual 1000 ppm stock standards
 
[ Care and Use ManUal ]
Waters IC-Pak C M/D Column
4
General considerations
•
Filter all aqueous buffers. Do not use turbid or cloudy buffers.
•
Do	not	reuse	aliquots	taken	from	the	sample	container;	dispose	of	
them.
•
Protect	the	column	from	vibration,	mechanical	shock,	and	rapid	
changes in pressure, flow rate or mobile phase composition. Any 
thermal, physical or chemical shock (such as changing mobile phases 
rapidly) can cause a loss of efficiency.
•
When	using	water,	treat	it	with	a	water	system	capable	of	delivering			
18 megohm water. Neither deionized water nor HPLC-grade bottled 
water are acceptable because they may contain organic compounds 
which alter column selectivity.
•
Highly	concentrated	samples	(greater	than	10	ppm	per	ion)	can	
overload the packing material, resulting in poor peak shape. Dilute 
the sample before injection. When analyzing an unknown, start with a 
1:100 dilution.
•
DO NOT inject concentrated samples directly into the mobile phase. 
Direct injection may cause precipitation of the salts in the sample. 
Dissolve (or dilute) samples in an appropriate volume of the mobile 
phase first. If you must use other solvents, watch for precipitation upon 
injection into the eluent. Always filter samples before use.
b. Efficiency Testing
Perform on efficiency test before attempting the first analysis. 
Run the test sample using the calibration standards detailed in the 
following pages. Record the retention time and the settings used
There are four parts to performing the efficiency test: 
1. Preparing the mobile phase 
2. Preparing the stock calibration standards
3. Running the calibration (working) standard
4. Calculating column efficiency
If you experience problems during normal operation, repeat the 
conditions for the initial efficiency test and compare the results. 
Differences may indicate a problem with the column.
Preparing Mobile Phase
The IC-Pak C M/D column uses 0.1 mM EDTA/3.0 mM HNO
3
as a
mobile phase. See phase Section III. a, Preparing the Mobile Phase, 
for procedures for mobile phase preparation.
Guidelines for the Calibration Standard
Standard concentrates are available from most major chemical 
suppliers. Use reagent-grade or analytical standard-grade 
solutions. Several anionic species can cause the precipitation of 
alkali and alkaline earth metals. Consult solubility tables to avoid 
these species. Avoid hygroscopic salts. Select the highest purity 
salt available. Certain atomic absorption standards are made 
from ammonium salts and must not be used to prepare calibration 
standard.
Preparing Individual 1000 ppm Stock Standards
1. Weigh the specified amount of salt (refer to Table 1).
2. Add the salt to a plastic 1 liter volumetric flask.
3. Fill the flask to the mark with 18 megohm water.
4. Store the stock solutions in clean plasticware at room temperature for
up to 6 months.
Table 1: Salt Weight for Stock Solution Preparation
The following equation is an example of how these weights were 
determined: 
1 g K
+
/L x 74.553 g KCI/39.100 g K
+
x 1 L = 1.9067 g KCI
Cation
Compound
Weight (g)
Li
+
(Lithium)
UOH • H
2
0
6.0476
No
+
(Sodium)
NaCI
2.5421
NH
4
+
(Ammonium)
NH
4
CI
2.9640
K
+
(Potassium)
KCI
1.9067
Mg
2
+
(Magnesium)
Mg(NO
3
)
2
• 6H
2
0
10.5466
Ca
2
+
(Calcium)
Ca(N0
3
)
2
• 4H
2
0
5.8919
Sr
2
+
(Strontium)
Sr(NO
3
)
2
• 4H
2
0
3.2377
Ba
2
+
(Barium)
BaCl
2
• 2H
2
0
1.7786