Ultrahydrogel Columns

2

II. InstallatIon

a. Solvents

Ultrahydrogel columns are packed in distilled water for shipping.
Table 1 shows the maximum solvent replace¬ment flow rate for each
Ultrahydrogel column. Perform solvent changeover at a lower flow
rate than that shown in Table 2 for analysis.

CAUTION: ULTRAHYDROGEL COLUMNS ARE FLOW (PRESSURE) SENSI-
TIVE AND EXTREME CARE MUST BE TAKEN NOT TO EXCEED MAXIMUM
FLOW RATE ON START-UP.

b. Column

Table 1. Flow Rates for Replacing Solvents

NOTE: CHANGE SOLVENTS AS SELDOM AS POSSIBLE. FREQUENT
SOLVENT EXCHANGES ACCELERATE THE DEGRADATION OF COLUMN
EFFICIENCY.

c. Organic Solvents

Ultrahydrogel columns are physically and chemically stable in
ordinary water-soluble organic solvent mix¬tures. Aqueous solutions
of methanol, ethanol, acetonitrile, formic acid, and dimethylsulfoxide
also provide dependable separations. Organic solvent concentration
should not exceed 20%. It is possible however, to use organic sol-
vents up to 50% concentration by gradual introduction via gradient
method.

d. Aqueous Salt and Buffer Solution

Although some non-ionic samples can be analyzed in distilled water,
most water soluble polymers require aqueous salt or buffer solutions.

Some typical aqueous salt solutions:

•

Sodium sulfate

•

Sodium acetate

•

Sodium dihydrogenphosphate

•

Ammonium acetate

•

Ammonium formate

Compatible buffer solutions:

•

Phosphate

•

Tris phosphate

•

Tris hydrochloric acid

•

Citrate

•

Tris acetate

•

Acetate

e. Flow Rate

For best resolution and maximum column life the flow rate should
not exceed the maximum flow rate listed below. Use Table 2 below to
determine recommended flow rates and maximum pressures.

Table 2. Recommended Flow Rates

Column

Column Size

Max. Flow Rate During

Solvent Change

Ultrahydrogel (all pore sizes)

7.8 mm i.d. x 30 cm

0.3 ml/min

Ultrahydrogel DP

7.8 mm i.d. x 30 cm

0.3 ml/min

Ultrahydrogel DNA

7.8 mm i.d. x 30 cm

0.3 ml/min

Ultrahydrogel Linear

7.8 mm i.d. x 30 cm

0.3 ml/min

Column

Typical Operating Flow

Rate (mL/min)

Maximum Flow Rate

(mL/min)

Ultrahydrogel 120

0.5-0.8

1.0

Ultrahydrogel 250

0.5-0.8

1.0

Ultrahydrogel 500

0.3-0.6

1.0

Ultrahydrogel 1000

0.3-0.6

1.0

Ultrahydrogel 2000

0.3-0.6

1.0

Ultrahydrogel Linear

0.3-0.6

1.0

Ultrahydrogel DP

0.5-0.8

1.0