Care and use manual, D. slipfree connectors, Minimization of band spreading – Waters XTerra and XTerra Prep Columns User Manual

Page 3: F. measuring system bandspread volume, G. measuring gradient delay volume

Advertising
background image

[ Care and Use ManUal ]

XTerra Columns

3

d. SLIPFREE Connectors

A SLIPFREE Connector guarantees a void-free connection because
it pushes the tubing into the endfitting. This design comes
installed on the tubing. Fingertight to 10,000 psi – never needs
wrenches. Readjusts to all column endfittings. Compatible with all
commercially available endfittings. Unique design separates tube-
holding function from sealing function.

Minimization of band spreading

The following figure shows the influence of tubing internal
diameter on system band spreading and peak shape. As can be
seen, the larger tubing diameter causes excessive peak broadening
and lower sensitivity.

Figure 5: Effect of Connecting Tubing on System

f. Measuring System Bandspread Volume

1. Disconnect column from system and replace with a zero dead

volume union

2. Flow rate 1 mL/min. This should be performed on a single

wavelength detector (not a PDA/DAD)

3. Dilute a test mix in mobile phase to give a detector sensitiv-

ity 0.5-1.0 AUFS (can use the system start up test mix
which contains uracil, ethyl and propyl parabens; Waters
P/N WAT034544.)

4. Inject 2 to 5 µL of this solution.

5. Using 5 sigma method measure the peak width at 4.4% of peak

height:

Bandspread (µL) = PW (seconds)/60 (see Figure 6)

Typical LC system should be 100 µL ± 30 µL

Microbore (2.1mm i.d. and smaller) system should be no
greater than 20-40 µL

Figure 6: Determination of System Bandspread Volume using the
5-Sigma Method

g. Measuring Gradient Delay Volume

1. Replace the column with a zero dead volume union.

2. Determine the gradient-delay or dwell volume for your system

by performing the following test. Prepare eluent A (pure solvent,
such as methanol) and eluent B (solvent plus sample, such as
5.6 mg/mL propylparaben in methanol).

3. Equilibrate the system with eluent A until a stable baseline is

achieved. Switch to 100% eluent B and record the half height of
the step. Refer to Figure 7 for an illustration. The dwell volume
should be less than 1 mL. If this is not the case, see section on
System Modifications to reduce your system volume.

System Volume

4.4 %h

5

Waters Part Numbers for SLIPFREE Pre-assembled Tubing

Tubing Length

0.005” i.d.

0.010” i.d.

0.020” i.d.

Single SLIPFREE

®

6 cm

PSL 618000

PSL 618006

PSL 618012

Single SLIPFREE

®

10 cm

PSL 618002

PSL 618008

PSL 618014

Single SLIPFREE

®

20 cm

PSL 618004

PSL 618010

PSL 618016

Double SLIPFREE

®

6 cm

PSL 618001

PSL 618007

PSL 618013

Double SLIPFREE

®

10 cm

PSL 618003

PSL 618009

PSL 618015

Double SLIPFREE

®

20 cm

PSL 618005

PSL 618011

PSL 618017

Figure 4: Single and Double SLIPFREE Connectors

Diluted/Distorted Sample Band

0.005 inches

0.020 inches

0.040 inches

Advertising