Care and use manual – Waters XBridge Columns User Manual

Page 8

Advertising
background image

[ Care and Use ManUal ]

XBridge

Columns

8

5. Measure the peak width at 4.4% of peak height (5-sigma method):

5-sigma Bandspreading (µL) = Peak Width (min) x
Flow Rate (mL/min) x (1000 µL/1 mL)

System Variance (µL

2

) = (5-sigma bandspreading)

2

/ 25


Figure 7:
Determination of System Bandspreading Volume Using 5-Sigma
Method

In a typical HPLC system, the Bandspreading Volume should be no
greater than 100 µL ± 30 µL (or Variance of 400µL

2

± 36µL

2

).

In a microbore (2.1mm i.d.) system, the Bandspreading Volume should be no
greater than 20 to 40 µL (or Variance no greater than 16µL

2

to 64µL

2

).

c. Measuring Gradient Delay Volume (or Dwell Volume)

For successful gradient-method transfers the gradient delay volumes should be
measured using the same method on both HPLC systems. The procedure below
describes a method for determining the gradient delay volumes.

1. Replace the column with a zero dead volume union.

2. Prepare mobile phase A (pure solvent, such as methanol) and mobile phase

B (mobile phase A with a UV absorbing sample, such as (v/v) 0.1% acetone
in methanol).

3. Equilibrate the system with mobile phase A until a stable baseline is

achieved.

4. Set the detector wavelength to the absorbance maximum of the probe (265

nm for acetone).

5. Program a 0-100% B linear gradient in 10 min at 2 mL/min (the exact

conditions are not critical; just make sure the gradient volume is at least 20
mL) with a hold at 100% B.

6. Determine the dwell time by first locating the time at the midpoint of the

formed gradient (t

1/2

) (half the vertical distance between the initial and final

isocratic segments as shown in Figure 8).

7. Subtract half the gradient time (

1

/

2

t

g

) (10 min/2 = 5 min in this

example) from the gradient midpoint (t

1/2

) to obtain the dwell time (t

D

).

8. Convert the dwell time (t

D

) to the dwell volume (V

D

) by multiplying

by the flow rate (F).

Dwell Volume V

D

= (t

1/2

1

/

2

t

g

) x F

For fast gradient methods, the gradient delay volume (or dwell volume)
should be less than 1 mL. If the gradient delay volume is greater than 1
mL, see System Modification Recommendations section on how to reduce
system volume.

VII. addItIonal InformatIon
a. Use of Narrow-Bore (3.0 mm i.d.) Columns

This section describes how to minimize extra column effects and provides
guidelines on maximizing the performance of a narrow-bore column in an
HPLC system. A 3.0 mm i.d. narrow-bore column usually requires no sys-
tem modifications. A 2.1 mm i.d. column, however, requires modifications
to the HPLC system in order to eliminate excessive system bandspreading
volume. Without proper system modifications, excessive system band-
spreading volume causes peak broadening and has a large impact on peak
width as peak volume decreases.

System Volume

4.4 %h

5

Time

1/2 Vertical

Distance

1.0

0.8

0.6

0.4

Au

0.2

0.0

t

1/2

Figure 8: Determination of Gradient Delay Volume

Advertising