[ method guidelines ]

Figure 5: Analysis of a Digested 25mer Phosphorothioate Oligonucleotide


HPLC system:

Waters BioAlliance

2796, PDA Detector with micro UV cell

Sample:

Detritylated 25mer phosphorothioate oligonucleotide mix

(CTC TCG CAC CCA TCT CTC TCC TTC T) digested with 3’ exonuclease

Column:

XBridge

OST C

18

, 2.5 µm (2.1 x 50 mm)

Mobile phase:

A: 15 mM TEA with 400 mM HFIP

B: methanol

Flow rate:

0.2 mL/min

Column Temp.:

60 ˚C

Gradient delay:

0 mL

Gradient:

15 to 20% B in 20 minutes (0.25% methanol per minute)

Detection:

260 nm, 2 scans per second

Peptide nucleic acids (PNA) can also be analyzed using XBridge

OST C

18

columns. The ion-pairing system recommended for analysis of PNA is
similar to those used for peptide analysis (0.1% Trifluoroacetic Acid
or Formic Acid).

UV detection of eluted oligonucleotide peaks is often performed at
260 nm. Injection of 50 pmol of detritylated oligonucleotide sample
on a 2.1 x 50mm XBridge

OST C

18

column yields relatively abundant

peaks. Limits of quantitation (LOQ) vary with the type of oligonucleotide,
LC system and detector; LOQ generated on Waters 2996 PDA detector
equipped with micro UV cell is approximately 1 pmol (2.1 x 50mm
XBridge

OST C

18

). The Limit of detection (LOD) estimate is shown in

Figure 6.

Figure 6: Analysis of a 20mer Oligodeoxythymidine Crude Synthesis Mixture

HPLC system:

Waters BioAlliance

2796, PDA Detector with micro UV cell

Sample:

~600 pmol of a detritylated 20mer, ~18 pmol of 19mer, ~4.5 pmol

of 17mer was injected on column.

Column:

XBridge

OST C

18

, 2.5 µm (2.1 x 50mm)

Mobile phase:

A: 0.1 M TEAA,

B: Acetonitrile / 0.1M TEAA, 20/80 (v/v)

Flow rate:

0.2 mL/min

Column Temp:

60 ˚C

Gradient Delay:

0 mL

Gradient:

35 to 50% B in 30 minutes (7-10% acetonitrile)

Detection:

260 nm, 2 scans per second

Vii. puriFiCation Considerations

XBridge

OST C

18

columns are designed for laboratory scale oligo-

nucleotide purifications and analyses. Sufficient amount of isolated
material suitable for molecular biology and other experiments can
be prepared in a single injection. For example, a 4.6 X 50 mm XBridge

OST C

18

column can suitably purify approximately 20-200 nmoles of

sample in a single injection. It is important to understand that column
overloading results in a peak broadening and that some earlier eluting
impurities may co-elute with the component of interest. With a proper
heart-cutting technique, a good purity of the target oligonucleotide
can be obtained without significant yield sacrifice (Figure 7).

Chromatographers frequently develop a separation on the analytical
scale before moving to preparative work. The steps required to opti-
mize the analytical separation involve:

1) Selecting the appropriate column packing material and mobile phase.

2) Determining the optimal flow rate, gradient during and separation

temperature.

3) Determining the amount of material that can be satisfactorily

loaded and separated on the analytical scale column.