Metrohm 781 pH/Ion Meter: Operation via RS-232 User Manual

Page 12

Advertising
background image

1.4 Error messages

10

780/781 pH/Ion Meter - Operation via RS232

&Mode

Object Description

Input

range

&Mode

Mode

$G, $S, $H, $C

:

.

Select

Mode selection

pH, U, T, Conc

.

pH

Measuring mode pH

│ ├

.Cal

Start or stop calibration

$G, $S

│ ├

.ElTest

Start or stop electrode test

$G, $S

│ ├

.MeasPara

Measuring parameters

│ │

.ElectrodeId

Electrode identification

12 ASCII char.

│ │

.Drift

Drift for meas.value acquisition [/min] 0.001...0.050...9.999, OFF

│ │

.Temperature

Measuring temperature [°C]

-999.9...25.0...999.9

│ │

.MethodId

Method name

8 ASCII char. (read only)

│ │

.Delta Delta

measurement

│ │

.Status

Status of delta measurement

ON, OFF

│ │

.Reference

Reference for delta measurement

-19.999...0.000...19.999

│ │

.Stirrer

│ │

.Status Status

of

stirrer

ON,

OFF, control

│ │

.Rate Stirring

rate

1...5...15

│ │

.PreStirTime Pause before stirring [s]

0...99999

│ │

.StirTime

Stirring time [s]

0...99999

│ │

.PostStirTime Pause after stirring [s]

0...99999

│ ├

.CalPara

Calibration parameters

│ │

.CalTemp

Calibration temperature [°C]

0.0...25.0...99.0

│ │

.Drift

Drift f. meas. val. acquisition [mV/min] 0.1...0.5...9.9

│ │

.Report

Automatic printout of calibration report full, short, OFF

│ │

.CalInterval

Calibration interval [h]

1...999, OFF

│ │

.Buffer

│ │

.Number

Number of calibration buffers

1...2...9

│ │

.Type

Type of calibration buffer set

Metrohm, NIST, DIN, Fisher, Fluka-BS,

│ │

Mettler, Merck Tit., Merck Cer., Beckman,

Radiometer, Baker, Hamilton, Precisa,

│ │

special, own, mixed

│ │

.Special Special

buffers

│ │

.1

│ │

.Val

pH value for buffer 1

-19.999...7.000...19.999

│ │

.2

│ │

.Val

pH value for buffer 2

-19.999...4.000...19.999

│ │

.3

│ │

.Val

pH value for buffer 3

-19.999...7.000...19.999

│ │

up to 9

│ │

.Own

Definition of own buffers

│ │

.1

Buffer 1

│ │

.1

Temperature 1 = 0 °C

│ │

.Val pH value for buffer 1 at 0 °C

-19.999...19.999, OFF

│ │

up to 20 Temperature up to 95 °C in 5 °C steps

│ │

up to 5

Up to 5 own buffers

│ │

.Mixed

Definition of mixed buffers

│ │

.1

Buffer 1

│ │

.Select Selection of buffer 1 Met4, Met7, Met9, NIST1, NIST4,

NIST7,

│ │

NIST9, NIST13, DIN1, DIN3, DIN4, DIN7,

│ │

DIN9, DIN12, Fis2, Fis4, Fis7, Fis10,

│ │

FBS4, FBS7, FBS9, MT2, MT4, MT7, MT9,

│ │

MT11, Mer1, Mer2, Mer3, Mer4, Mer4.66,

│ │

Mer5, Mer6, Mer6.88, Mer7, Mer8, Mer9,

│ │

Mer9.22, Mer10, Mer11, Mer12, Mer13,

│ │

MerC4.01, MerC7.00, MerC9.00, MerC10.0,

Advertising