Software-handshake, swline, Ph/ion meter as receiver, Ph/ion meter as sender – Metrohm 781 pH/Ion Meter: Operation via RS-232 User Manual

Page 25

Advertising
background image

3 RS232 interface properties

780/781 pH/Ion Meter - Operation via RS232

23

Software-Handshake, SWline

Handshake input ports on the pH/Ion Meter (CTS) are not checked.
Handshake output ports (DTR, RTS) are set by the pH/Ion Meter.
The pH/Ion Meter has an input buffer which can accept up to 512

characters.

pH/Ion Meter as Receiver :

pH/Ion Meter as Sender:

pH/Ion Meter transmission can be stopped by external instruments with

XOFF. After XOFF is received the pH/Ion Meter completes sending the

line already started. If data output is disabled for more than 6 s by

XOFF, E43 appears in the display.

pH/Ion Meter

external device

Data output

Data input

max. 128

characters

384 characters

XOFF

XON

L

F

RxD

TxD

Time

pH/Ion Meter

external device

XOFF

XON

Data output

disabled

Data output

enabled

Time

Data output

Data input

RxD

TxD

2

nd

line

L

F

1

st

line

L

F

Advertising