2 triggers, Triggers – Metrohm 781 pH/Ion Meter: Operation via RS-232 User Manual

Page 5

Advertising
background image

1 General rules

780/781 pH/Ion Meter - Operation via RS232

3

1.2 Triggers

Triggers initiate an action on the pH/Ion Meter. They are marked by the

introductory symbol $.
The following triggers are possible:
$G

Go

Starts processes, for ex. starting the mode run or

setting the RS232 interface parameters

$S Stop Stops

processes

$Q

Query

Queries all information from the current node in the

tree forward up to and including the values

$Q.P Path

Queries the path from the root of the tree up to the

current node

$D

Detail-Info Queries the detailed status information

$U

qUit

Aborts the data flow of the instrument, for example,

after $Q

The triggers $G and $S are linked to particular objects, see section 1.1.
All other triggers can be used at any time and at all locations on the

object tree.
Examples:
Querying the value of the baud rate: &Config.RSSet.Baud $Q
Querying all values of the node "RSSet": &Config.RSSet $Q
Querying the path of the node "RSSet": &Config.RSSet $Q.P
Start of pH calibration: &Mode.pH.Cal $G
Querying the detailed status: $D