Metrohm 781 pH/Ion Meter: Operation via RS-232 User Manual

Page 14

Advertising
background image

1.4 Error messages

12

780/781 pH/Ion Meter - Operation via RS232

│ │

.ElParam Electrode

parameters

│ │ │ │

.StreamPot Streaming potential [mV]

-999.9...4.0...999.9

│ │ │ │

.Drift Drift

[mV/min]

0.1...3.0...9.9

│ │ │ │

.MinSlope Minimum slope [%]

0.1...95.00...999.9

│ │ │ │

.MaxSlope Maximum slope [%]

0.1...103.0...999.9

│ │ │ │

.Response Response time [s]

0...60...9999

│ │

.MinUas

Lower limit Uoff [mV]

-2200.0...-15.0...2200.0

│ │

.MaxUas

Upper limit Uoff [mV]

-2200.0...15.0...2200.0

│ │

.Temperature

Temperature for electrode test [°C]

0.0...25.0...99.9

│ │

.Report

Report

short, full, line, OFF

.U

Measuring mode U (potential)

│ ├

.MeasPara

Measuring parameters

│ │

.ElectrodeId

Electrode identification

12 ASCII char.

│ │

.Drift

U drift f. meas. val. Acquisit. [mV/min] 0.1...1.0...999.9, OFF

│ │

.MethodId

Method name

8 ASCII char. (read only)

│ │

.Delta Delta

measurement

│ │

.Status

Status of delta measurement

ON, OFF

│ │

.Reference

Reference for delta meas. [mV]

-2200.0...0.0...2200.0

│ │

.Stirrer

│ │

.Status Status

of

stirrer

ON,

OFF, control

│ │

.Rate Stirring

rate

1...5...15

│ │

.PreStirTime Pause before stirring [s]

0...99999

│ │

.StirTime

Stirring time [s]

0...99999

│ │

.PostStirTime Pause after stirring [s]

0...99999

│ │

.Limits

│ │

.Status Status

limits

U

ON,

OFF

│ │

.UpperLim

Upper limit U [mV]

-2200.0...1000.0...2200.0

│ │

.UHystereses

Upper hysteresis U [mV]

-2200.0...2.0...2200.0

│ │

.LowerLim Lower

limit

U [mV]

-2200.0...-1000.0...2200.0

│ │

.LHystereses

Lower hysteresis U [mV]

-2200.0...2.0...2200.0

│ ├

. PlotPara

Parameters for live plot on a printer

│ │

.LeftMargin

Left margin of voltage scale [mV]

-1999...-1000...1999

│ │

.RightMargin Right

margin

of

voltage scale [mV]

-1999...1000...1999

│ └

.Presel

Preselections

.IReq

Request of identifications after start id1, id1 & id2, OFF

.T

Measuring mode T (temperature)

│ ├

.MeasPara

Measuring parameters

│ │

.ElectrodeId

Electrode identification

12 ASCII char.

│ │

.Drift

Drift for meas. val. acquisition [°C/min] 0.5...1.0...999.9, OFF

│ │

.MethodId

Method name

8 ASCII char. (read only)

│ │

.Delta Delta

measurement

│ │

.Status

Status of delta measurement

ON, OFF

│ │

.Reference

Reference for delta measurement [°C] -999.9...0.0...999.9

│ │

.Stirrer

│ │

.Status Status

of

stirrer

ON,

OFF, control

│ │

.Rate Stirring

rate

1...5...15

│ │

.PreStirTime Pause before stirring [s]

0...99999

│ │

.StirTime

Stirring time [s]

0...99999

│ │

.PostStirTime Pause after stirring [s]

0...99999

│ │

.Limits

│ │

.Status

Status limits T

ON, OFF

│ │

.UpperLim

Upper limit T [°C]

-999.9...100.0...999.9

│ │

.UHystereses

Upper hysteresis T [°C]

-999.9...0.2...999.9

│ │

.LowerLim

Lower limit T [°C]

-999.9...0.0...999.9

│ │

.LHystereses

Lower hysteresis T [°C]

-999.9...0.2...999.9

│ ├

.PlotPara

Parameters for live plot on a printer

Advertising
This manual is related to the following products: