3 RS232 interface properties

780/781 pH/Ion Meter - Operation via RS232

21

3 RS232 interface properties

3.1 Data

transfer

protocol


The pH/Ion Meter is configured as DTE (Data Terminal Equipment).
The RS232 interface has the following technical specifications:
• Data interface according to the RS232C standard, adjustable

transfer parameters, see page 16.

• Max. line length:

80 characters (incl. CR LF)

• Control characters:

C

R

(ASCII DEC 13)

L

F

(ASCII DEC 10)

XON (ASCII DEC 17)

XOFF (ASCII DEC 19)

• Cable length:

max. approx. 15 m

Start

7 or 8 Data Bit

Parity Bit

1 or 2 Stop Bit

Only a shielded data cable (for example, Metrohm 6.2125.110 DB-

9/DB-25 or 6.2134.100 DB-9/DB-9) may be used to couple the pH/Ion

Meter with foreign devices. The cable shield must be properly grounded

on both instruments (pay attention to current loops; always ground in a

star-head formation). Only connectors with sufficient shielding may be

used