Metrohm 781 pH/Ion Meter: Operation via RS-232 User Manual

Page 16

Advertising
background image

1.4 Error messages

14

780/781 pH/Ion Meter - Operation via RS232

.CMax

Maximal concentration of standard

1.0E-30...1.0...1.0E+30

.VInit

Total initial volume (TISAB etc.) [mL] 0.001...100.0...999.9

.NoExUnit Number of exchange units

1...5

.1

.Conc Conc. of standard in exchange unit 1 1.0E-30...100.0...1.0E+30

.ExV Volume of exchange unit 1 [mL]

1, 5, 10, 20, 50

up to 5 Up to 5 exchange units

│ ├

.LimitsConc

│ │

.Status Status

limits concentration

ON, OFF

│ │

.UpperLim Upper

limit

conc

-1.00E+30...1.00E+30

│ │

.UHystereses Upper

hysteresis

conc

-1.00E+30...2.00...1.00E+30

│ │

.LowerLim Lower

limit

conc

-1.00E+30...0.00...1.00E+30

│ │

.LHystereses Lower hysteresis conc

-1.00E+30...2.00...1.00E+30

│ ├

.LimitsT

│ │

.Status

Status limits T

ON, OFF

│ │

.UpperLim

Upper limit T [°C]

-999.9...100.0...999.9

│ │

.UHystereses Upper hysteresis T [°C]

-999.9...0.2...999.9

│ │

.LowerLim

Lower limit T [°C]

-999.9...0.0...999.9

│ │

.LHystereses Lower hysteresis T [°C]

-999.9...0.2...999.9

│ └

.PlotPara

Parameters for live plot on a printer

.LeftMargin

Left margin of Conc scale

-1.0E+30...0.0...1.0E+30

│ ├

.RightMargin Right

margin

of Conc scale

-1.0E+30...1.0E+30

.TLeftMarg

Left margin of temperature scale [°C] -999...20...999

.TRightMarg Right margin of temperature scale [°C] -999...30...999

.StdAdd

Parameters for standard addition $G, $S

.Type Type of standard addition

add, sub

.Conc

Concentration of standard solution

1.0E-30...1.0...1.0E+30

.Report

Selection of report type

short, full, line, OFF

.Add

Type of standard addition

manual, auto dos, auto

.DeltaU For

automatic

additions [mV]

1...10...999

.DosRate

Dosing rate for addition

slow, medium, fast

.NumberAdd

Number of additions

1...19

.Bur

Volume of exchange unit [mL]

1, 5, 10, 20, 50

.Increment Size

of

increment

│ │

.1

First increment

│ │

.Val Volume

[mL]

1.0E-30...0.1...99.999

│ │

up to 19

Up to 19 additions

.StopV

Stop volume [mL]

0.0...99.99...9999.99

.SmplAdd

Parameters for sample addition

$G, $S

.Type

Type of sample addition

add, sub

.Conc

Concentration of standard solution

1.0E-30...1.0...1.0E+30

.Report

Selection of report type

short, full, line, OFF

.Add

Type of sample addition

manual, auto dos, auto

.DeltaU For

automatic

additions [mV]

1...10...999

.DosRate

Dosing rate for addition

slow, medium, fast

.NumberAdd

Number of additions

1...19

.Bur

Volume of exchange unit [mL]

1, 5, 10, 20, 50

.Increment Size

of

increment

│ │

.1

First increment

│ │

.Val Volume

[mL]

1.0E-30...0.1...99.999

│ │

up to 19

Up to 19 additions

.StopV

Stop volume [mL]

0.0...99.99...9999.99

.Presel

Preselections

.IReq

Request of identifications after start id1, id1 & id2, OFF

.SReq

Request of Sample size after start

value, unit, all, OFF

.ActivatePulse

Output of a pulse

ON, OFF

Advertising