3 pin assignment, Pin assignment – Metrohm 781 pH/Ion Meter: Operation via RS-232 User Manual

Page 27

Advertising
background image

3 RS232 interface properties

780/781 pH/Ion Meter - Operation via RS232

25

3.3 Pin

assignment

RS232C Interface

Transmitted Data (TxD).
If no data are transmitted, the line is held in the “ON”
condition. Data will only be sent when CTS is in the ”ON”
condition.

Received Data (RxD)



Request to Send (RTS)
ON condition: pH/Ion Meter is ready to send data.


Clear to Send (CTS)
ON condition: Remote station is ready to receive data.



Signal ground (GND)



Data Terminal Ready (DTR)
ON condition: Instrument is ready to receive data.

Protective earthing
Direct connection from cable plug to the protective ground of the

instrument.

Polarity allocation of the signals
-

Data lines (TxD, RxD)

voltage negative (<-3 V): signal state ”ON”

voltage positive (>+3 V): signal state ”ZERO”

-

control or message lines (CTS, RTS, DTR)

voltage negative (<-3 V): OFF state

voltage positive (>+3 V): ON state

In the transitional range from +3 V to -3 V the signal state is undefined.

Driver according to EIA RS232C specification.
Receiver according to EIA RS232C specification.

Pin 4

Data Terminal Ready

Pin 5

Signal Ground

Pin 8

Clear To Send

Pin 7

Request To Send

Pin 2

Received Data

Pin 3

Transmitted Data

Advertising