Metrohm NIRS XDS Transmission OptiProbe Analyzer User Manual
Page 46

44
▪▪▪▪▪▪▪
This dialog box is displayed.
Next a sample scan is taken (on the
reference) and this box is displayed. This
takes about 20-30 seconds.
Wavelength Linearization is performed in two
sections, each of which takes about 45
seconds.
The user must confirm the linearization to
proceed. A dialog box will appear, and the
user should click “Yes” to proceed to the next
step.
Following Wavelength Linearization, the user
is asked to select a standard file. This is
provided on the mini-CD in the standard box.
(It may be manually copied on the computer
in the Vision directory, if desired.)
Select this file and click on “Open”.
Vision acquires a new reference scan.
Vision requests that the XC-1300 WST3Wxxx
standard be placed into the calibration
fixture.