4 fuse replacement, Fuse replacement – Metrohm NIRS XDS Transmission OptiProbe Analyzer User Manual
Page 75
▪▪▪▪▪▪▪
73
Perform tests as specified in the prior sections of
this manual. These normally include the
following:
•
Wavelength Linearization
•
Performance Test
•
Reference Standardization
(if applicable to this model)
•
Instrument Calibration
The spare lamp may come with generic
instructions for instrument testing. Use the
instructions in this manual to test the instrument,
as this manual overrides all instructions packed
with the lamp.
Immediately contact your Metrohm distributor to
order a new spare lamp.
The part number is 6801 4801, model number
XA-3000. It is wise to keep several spare lamps
on hand.
Do not accept lamps from any other source. They
will not work in your Metrohm instrument.
8.4
Fuse Replacement
Fuse Replacement is an unusual event, and usually is caused by some electrical fault. The electrical
fault should always be investigated and repaired before fuse replacement. Once the fault is found
and corrected, this procedure should be followed:
Turn off power at the power switch located on
the back panel.