4 replacing parts of the suppressor – Metrohm 881 Compact IC pro – Anion – MCS User Manual

Page 101

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Operation and maintenance

881 Compact IC pro – Anion – MCS

■■■■■■■■

91

5 Inserting the rotor

Caution

An incorrectly inserted rotor can be destroyed during start-up.

■

Insert the rotor (42-3) into the housing (42-4) in such a way that
the tubing connections on the rear of rotor fit into the corre-
sponding recesses inside the housing and one of the three holes
of the rotor is visible from below in the slot of the housing (42-5).

■

If the rotor is correctly inserted, its sealing surface will be approx.
4 mm within the housing. If this is not the case, the rotor must be
moved into the right position from below using a sharp object.

6 Cleaning the connecting piece

■

Clean the surface of the connecting piece (42-2) with ethanol
using a lint-free cloth.

7 Inserting the connecting piece

■

Insert the connecting piece (42-2) into the housing in such a way
that the connector 1 is on top and the three pins of the connect-
ing piece fit into the corresponding recesses on the housing.

■

Reattach the union nut (42-1) onto the housing and tighten by
hand (do not use a tool).

8 Connecting and conditioning the suppressor

■

Reconnect the suppressor to the IC system.

■

Before switching the suppressor over for the first time, rinse each
of the three suppressor units with solution for 5 minutes.

5.12.3.4

Replacing parts of the suppressor

In the following cases, it might be necessary to replace parts of the sup-
pressor:

■

Loss of suppression capacity which cannot be eliminated (reduced
phosphate sensitivity and/or significant rise in the baseline).

■

Blockage of the suppressor which cannot be eliminated (solutions can
no longer be pumped through the suppressor).

Both the rotor and the connecting piece can be replaced.