5 Operation and maintenance

881 Compact IC pro – Anion – MCS

■■■■■■■■

89

Regenerating the suppressor

1 Disconnecting the suppressor from the IC system

■

Disconnect the capillaries of the suppressor labeled with regener-
ant
and rinsing solution from the IC system.

2 Connecting the suppressor to the high pressure pump

■

Connect the inlet capillary for the regeneration solution (labeled
with regenerant) to the outlet of the high pressure pump with the
aid of a coupling (6.2744.040)

3 Regenerating the suppressor

■

Regenerate the first suppressor unit for about 15 minutes.

■

In the software, use the Step command to switch to the second
suppressor unit and regenerate this for about 15 minutes.

■

In the software, use the Step command to switch to the third
suppressor unit and regenerate this for about 15 minutes.

4 Rinsing the suppressor

After regeneration, the three suppressor units must be rinsed with
degassed ultra pure water for about 15 minutes.

■

Remove the inlet capillary for the regeneration solution (labeled
with regenerant) from the outlet of the high pressure pump.

■

Connect the inlet capillary for the rinsing solution (labeled with
rinsing solution) to the outlet of the high pressure pump with the
aid of a coupling (6.2744.040).

■

Rinse the first suppression unit with degassed ultra pure water for
about 15 minutes.

■

In the software, use the Step command to switch to the second
suppressor unit and rinse this for about 15 minutes.

■

In the software, use the Step command to switch to the third
suppressor unit and rinse this for about 15 minutes.

5 Connect the suppressor to the IC system

■

Reconnect the capillaries labeled regenerant and rinsing solution
to the IC system.