3 Installation

881 Compact IC pro – Anion – MCS

■■■■■■■■

13

■

Attach the CO

2

adsorption cartridge (6.2837.000) to the adsorp-

tion cartridge holder (6.2057.080) (see "Installing the adsorption
cartridges", page 57)
.

■

Prepare the H

2

O adsorption cartridge (6.2837.010) (see leaflet to

the H

2

O adsorption cartridge) and attach it to the adsorption car-

tridge holder as well (see Figure 30, page 56).

■

Plug the adapter (6.1808.190) onto the PVC tubing and connect
the two adsorption cartridges with one another (see Figure 30,
page 56)
.

■

Place the adsorption cartridge (6.2057.080) holder in the detector
chamber.

■

Connect the MCS air aspirating capillary (3-15) to the tip of the
CO

2

adsorption cartridge (6.2837.000).

8 Connecting the instrument

■

Connect the instrument to a computer (see Chapter 3.19.1, page
58)
on which the software MagIC Net™ is installed using the
USB cable (6.2151.020).

■

Connect the instrument to the mains supply (see Chapter 3.19.2,
page 58)
.

9 Initial start-up

(see Chapter 4.1, page 63)

■

Switch on the PC and start MagIC Net™.

■

Switch on the instrument.

■

Deaerate the high pressure pump (see Chapter 3.10.2, page
33)
.

■

Set contact pressure of the peristaltic pump (see "Set flow rate",
page 50)
.

■

Rinse the instrument without column with eluent for 5 minutes.

10 Installing guard and separation column

■

Remove the coupling (6.2744.040) between the column inlet
capillary and the capillary of the MSM labeled with in.

■

(Optional) Connect guard column (see Chapter 3.20, page 59)

– Fasten the guard column to the end of the column inlet

capillary (see leaflet to the guard column).

– Rinse the guard column with eluent for approx. 5 minutes.