3 installing the adsorption cartridges, Installing the adsorption cartridges, Figure 30 – Metrohm 881 Compact IC pro – Anion – MCS User Manual

Page 66: Adsorption cartridge holder

Advertising
background image

3.18 Metrohm CO

2

Suppressor (MCS)

■■■■■■■■■■■■■■■■■■■■■■

56

■■■■■■■■

881 Compact IC pro – Anion – MCS

3.18.3

Installing the adsorption cartridges

For effective CO

2

removal, the air sucked through the degassing cell

should be as low in CO

2

as possible. To achieve this, the air is aspirated

through a CO

2

adsorption cartridge (6.2837.000) (30-4)..

Moisture can block the CO

2

adsorption cartridge. In order to prevent this,

a H

2

O adsorption cartridge (6.2837.010) (30-7) is connected upstream.

9

3

8

7

3

1

2

3

3

4

5

6

Figure 30

Adsorption cartridge holder

1

PEEK pressure screw, short
(6.2744.070)

2

Luer coupling (6.2744.120)

3

Clips
For fastening the adsorption cartridges.

4

CO

2

adsorption cartridge (6.2837.000)

For removing the CO

2

from the aspirated air.

3-layer filled, blue-brown-gray.

5

Adapter (6.1808.190)
For connecting the H

2

O adsorption cartridge

and CO

2

adsorption cartridge.

6

PVC tubing
For connecting the H

2

O adsorption cartridge

and CO

2

adsorption cartridge.