3 Installation

930 Compact IC Flex Oven/SeS/Deg (2.930.2460)

■■■■■■■■

41

Connecting the sample degasser

CAUTION

If the sample degasser is not used, the input and output must be
sealed with threaded stoppers (6.2744.220).

1 Removing the threaded stoppers

Remove and keep the threaded stoppers (6.2744.220) from the input
and output of the sample degasser.

2 Connecting the output capillary

■

Push a long pressure screw over the loose end of the capillary
connected to port 1 of the injection valve.

■

Tighten the pressure screw at the output of the sample degasser
(labeled Out).
Try to ensure the shortest possible connections.

3 Connecting the input capillary

■

Push a long pressure screw over one end of the PTFE capillary
(6.1803.040) and tighten the pressure screw at the input of the
sample degasser (labeled In).

■

Guide the other end of the PTFE capillary (6.1803.040) out of the
instrument through a capillary feed-through and connect it to the
Sample Processor, if applicable.