5.9 Servicing the inline filter

■■■■■■■■■■■■■■■■■■■■■■

74

■■■■■■■■

930 Compact IC Flex Oven/SeS/Deg (2.930.2460)

Connecting the input for the high-pressure pump

1

■

Remove the stopper from the coupling. Tighten the coupling to
the pressure screw located on the pump head input capillary (1).

■

Reconnect the pump head output capillary to the pump head out-
put (2).

5.9

Servicing the inline filter

NOTE

You can find a video sequence for this task in the Multimedia Guide IC
Maintenance
or on the Internet at

http://www.metrohm.com/com/

Support/Video/ICGeneral.html?q=14

.

Maintenance interval

The filter has to be replaced at least every 3 months; the filter has to be
replaced more frequently if you are working with high backpressure.

Accessories

For this task you need the following accessories:

■

Adjustable wrench (6.2621.000) from the accessory kit: Vario/Flex
Basic (6.5000.000)

■

Tweezers

■

A new filter from the packaging (6.2821.130)